June 28, 2016, 715003956 Rev. C
Page 95
Figure D–12:
Front configuration with LC stack
Tip:
This figure shows a QDa Detector fitted with the external, source exhaust valve assembly
(serial number suffix "KAB"). To see the QDa Detector fitted with the internal, source exhaust
valve (serial number suffix "KAD"), see the figures on
(rear configuration) and
(front configuration).
D.7.3
Connecting the 2.5-L waste container
Perform the following procedure to connect to the 2.5-L waste container to QDa models fitted with
a diaphragm pump, and QDa models fitted with the rear drain bulkhead, including the following
models:
•
Performance QDa Detector (serial number suffix "KAB")
•
Standard QDa Detector (serial number suffix "KAB")
•
Standard QDa Detector (serial number suffix "KAD")
Note:
The Performance QDa Detector fitted with the internal, source exhaust valve (serial number
suffix "KAD") does not connect to the 2.5-L waste container.
Source exhaust
valve assembly
To laboratory exhaust
system
Exhaust-trap bottle
ACQUITY QDa
PTFE tubing (ensure
downward gradient)
PTFE tubing
Front exhaust
connection
Convoluted
liquid waste
tubing
Liquid waste
container
90-degree
elbow
Exhaust solenoid cable