June 28, 2016, 715003956 Rev. C
Page 54
Figure 3–19:
Inserting the source exhaust valve
4.
Ensure that the valve is firmly in place so that the panel is flush with the outer edge of the
enclosure.
5.
Insert the four retaining screws in each corner of the grip panel, and tighten them using the
5-mm Allen wrench.
6.
Reconnect the source exhaust by threading the tubing through the hose brackets alongside
the instrument and inserting the tubing into the elbow connector. See
internal, source exhaust valve (KAD)”
Notes:
•
Routing the exhaust tubing through the hose brackets positions the tubing at a
downward gradient. Doing so prevents the pooling of solvents.
•
To prevent the pooling of solvents in a front configuration, ensure that the source
exhaust tubing is positioned forward on a downward gradient.
Source exhaust
valve
Source exhaust tube elbow
connector
Alignment key and grip panel
keyway
Valve grip panel
Source valve
enclosure
Retaining
screws