June 28, 2016, 715003956 Rev. C
Page 22
2.3
Connecting the probe
Before starting the instrument, you connect the LC flow to the probe.
To connect the probe:
1.
Connect the LC flow or syringe pump to the probe assembly’s input, and secure the PEEK
fitting.
2.
Carefully insert the probe capillary into the inlet atop the source enclosure, and tighten the
probe fitting until it clicks.
Warning:
To avoid puncture injuries from the sharp probe capillary, use care when inserting
it into the source enclosure.
Notice:
• To avoid damaging the fragile probe capillary, where the instrument is situated above
eye level, disconnect the instrument’s external connections and place it on a bench top
or other surface so that you can exercise due caution when inserting the probe (see
).
Warning:
To avoid puncture injuries from the sharp probe capillary, use care when
inserting it into the source enclosure.
Notice:
To avoid damage to the fragile probe capillary, use care when inserting it into
the source enclosure.