Making signal connections
2-13
Connecting to a chart recorder or external analog data collection
device
To send an analog output signal from the detector to a chart recorder, make
the connections shown in the following table and figure.
Analog output connections to a chart recorder
Analog output connections to a chart recorder
Chart recorder connectors
2998 PDA detector (connector B)
Pen 1
+
Pin 1 signal out
+
(red)
Pen 1 –
Pin 2 signal out – (black)
Tip:
For this connection, do not use the cable shield.
Y2
Y1
+ –
–
+
Chart recorder connectors
2998 PDA detector connector B
Red
Black
1
2
3
4
5
6
7
8
9
10
+
−
+
−
+
−
+
−
Analog 1
Analog 1
Ground
Analog 2
Analog 2
Switch 1
Switch 1
Ground
Switch 2
Switch 2
Summary of Contents for 2998