2-8
Setting Up the Detector
The following table summarizes the inject start connections for different
system configurations.
Tip:
If the injector is an e2695 separations module running in Ethernet mode
or a 2707 autosampler, then the inject start cable should not be connected.
However, if the injector is an e2695 separations module running in IEEE
mode, then the inject start cable should be connected.
Connecting to a manual injector
If you are using a manual injector with your system, connect the signal cables
from the rear panel connector on the detector to the injector as shown in the
following table.
For information on injection trigger signals from a manual injector, see
page 2-15
.
Connecting to other instruments
This section describes signal connections between the detector’s rear panel
and these items:
•
Waters Alliance Separations Module
•
Waters 1500-series pump
Detector inject start connections
Inject start output source
Inject start input connection (on the
2998 PDA detector, connector A)
Waters Alliance Separations Module Inject Start + / –
Waters 717 Autosampler
Inject Start + / –
Waters manual injector, or
third-party manual injector or
autosampler
Inject Start + / –
Detector connections to a manual injector
2998 PDA detector (connector A)
Manual injector
Inject Start
+
(red)
One set of spade lug Inject Start
terminals
Inject Start – (black)