2-24
Setting Up the Detector
5. Reinstall the detector front left panel cover:
•
To prevent invalid subsequent chromatographic results, remove the
cuvette from the detector, and replace the empty holder after
running your cuvette measurements.
•
To maintain optimum system performance, replace the front left
panel cover before resuming normal operation of the detector.
6. Insert a reference cuvette containing the mobile phase standard, and
run a zero measurement.
7. Replace the reference cuvette with a cuvette containing your analyte
dissolved in mobile phase solvent, and run a sample measurement.
8. Use the storage, review, subtract and review, and replay functions to
analyze the data obtained.