8. List of Paremeters 

8.1 Condition data 

Menu 

Parameter 

Default value 

Protect 

level 

(*1) 

Settable Selections / Ranges 

Function / Remarks 

P.L 

Protect Level 

PL0 

 

PL0/PL1/PL2/PL3 

Selects  the  protect  level  for  preventing  incorrect  operation. 
The higher the protect level, the more limitations are imposed 
on a set parameter. 

PVH 

PH select 

PH 

PL0 

PH/VH/PVH 

Selects  the  type  (peak  hold,  valley  hold,  or  peak-valley  hold) 
that is activated when the PH function is enabled. 

RANG 

Input Range 

14 

PL1 

11/12/13/14 

Selects the input range. 

25 

22/23/24/25 

AVG 

Averaging times 

1

 

PL1 

1/2/4/8/10/20/40/80/100/200 

Selects the number of averaging times (setting sampling rate). 
By setting as the number of averaging times of inner sampling 
of 25 times/sec (40 ms), the panelmeter practically uses the 
average as input and acts with the period of “AVG  ×  40ms”.   
Display of main display and output of BCD are also 
synchronized with this setting sampling. 
*See “10. Timing chart” for relationship the averaging times 
and setting sampling. 

MAV 

Moving averaging   
times 

OFF 

PL0 

OFF/2/4/8/16/32 

Selects the number of moving averaging times. (Lower filtering 
effect OFF⇔2⇔4⇔8⇔16⇔32 Higher filtering effect) 

S.WD 

Step wide 

PL0 

1/2/5/0 

Selects the resolution of the least significant digit. (When it is 
set to “5”, the least significant digit indicates only “0” or “5”.) 

BLNK 

Display blank 

level 

OFF 

PL0 

OFF/B-3/B-2/B-1/ON 

Selects display brightness. (Bright OFF⇔b-3⇔b-2⇔b-1⇔ON 
Extinguished) 

DL.T 

Digital limitter 

type 

CUT 

PL0 

CUT/OVER 

Selects display in case of overrange. When CUT is selected, the 
set value of DLHI/DLLO is displayed; when OVER is selected, 
O.L/-O.L is displayed. 

BCD.L 

BCD output logic 

N.LOG 

PL1 

N.LOG/P.LOG 

Selects the BCD output logic (N: negative logic, P: positive 
logic).     
* Only when BCD outputs are provided.

 

B.  UP 

DZ backup 

OFF 

PL0 

OFF/ON 

Selects whether to backup the digital zero value when power is 
dicconnected. 

LINE 

Linearize 

CLR 

PL0 

CLR/OFF/ON 

Selects the enable (ON) /disable (OFF) of the linearize function 
and data clear (CLR). 

TR.T 

Tracking zero 
correction time 

00 

PL0 

00~99 

Sets  the  correction  time  of  the  tracking  zero  function.  Every 
“TR.T×  setting sampling period”, the correction will be done. 
In case of TR.T is “00”, the tracking function is disabled. 

TR.W 

Tracking zero 

correction 
width 

01 

PL0 

01~99 

Sets the correction width of the tracking zero function. 
* Not available when TR.T is 00. 

PON 

Power on delay 

OFF 

PL0 

OFF,1~30 

Sets the time (set point x 1 sec.) taken from when the power is 
turned on to the instant when measurement is actually started. 

 
8.2 Scaling data 

Menu 

Parameter 

Default 

value 

Protect level 

Settable Selections / Ranges 

Function / Remarks 

FSC 

Full-scale 

display value 

19999 

PL1 

-19999  ~  19999 

Sets the relationship between an input signal and display value. 

FIN 

Full-scale input 

value 

19999 

PL1 

-19999  ~  19999 

 

OFS 

Offset display 

value 

PL1 

-19999  ~  19999 

*See,  ”Example  of  setting  scaling  data”  in  the  section”4.7 

Setting Scaling Data”. 

OIN 

Offset input 

value 

PL1 

-19999  ~  19999 

 

DLHI 

Digital  limiter 

High 

19999 

PL1 

-19999  ~  19999 

Sets  higher  limit  of  displayable  range.    For  higher  input  over 

this setting, indications are not  refreshed and kept the setting 

value. 

DLLO 

Digital  limiter 

Low 

-19999 

PL1 

-19999  ~  19999 

Sets lower limit of displayable range.  For lower input below this 

setting,  indications  are  not  refreshed  and  kept  the  setting 

value. 

DEP 

Decimal point 

..... 

PL1 

( Arbitrarily  settable  at  each 

digit) 

Sets the display position of decimal-point. 

If  “.....”  is set, no decimal-point is displayed. 

 
(*1) When s

etting value of “P.L” is “0”,    all parameters of PL0 to PL3 can be displayed and set. 

When setting value of “P.L” is “1”,    parameters of PL0 can not be displayed and parameters of PL1 to PL3 can be displayed and set. 
When setting value of “P.L” is “2”,    parameters of PL0 or PL1 can not be displayed and parameters of PL2 or PL3 can be displayed and set. 
When setting valu

e of “P.L” is “3”,    parameters of PL0 to PL2 can not be displayed and parameters of PL3 can be displayed and set. 

  (*2) Because all parameters in the scaling data is PL1, 

when setting value of “P.L” is 2 or 3, scaling data setting mode can not be entered. 

(i.e.    “NET” is never displayed when “P.L” is set to 2 or 3.)   

Summary of Contents for A9000 Series

Page 1: ...12 22 or 23 range Input terminal HI input terminal of 13 or 14 24 or 25 range Input terminal LO input terminal Selecting input range can not be done only by connecting input signal to a suitable termi...

Page 2: ...vel 3 5 5V Input current 0 5mA 4 Parameter Settings 4 1 Components and Functions Before setting parameters remove the front panel by inserting a flat blade screwdriver into the ditch under the front p...

Page 3: ...item 5 Press the Mode key to move to peak hold type setting 6 Set up the peak hold type using the Increment key This setting example sets peak hold type to PH PH Holds the maximum display value PV Ho...

Page 4: ...Enter and Mode keys during measurement operation to enter the condition data setting mode 2 Press the Shift key to move to the scaling data setting mode 3 Press the Mode key to display the number of...

Page 5: ...the setting again The number of linearize points is up to 16 but a value 17 to 19 is also displayed during setup Note that if you set a value from 17 to 19 it is forced to set to 16 5 Other Functions...

Page 6: ...23 19 999mA 24 199 99mA 0 1 3A 25 1999 9mA Common specifications A D conversion conversion method Input circuit Single ended Setting sampling rate 25times sec max Overrange warning For an input signa...

Page 7: ...ble ON disable OFF of the linearize function and data clear CLR TR T Tracking zero correction time 00 PL0 00 99 Sets the correction time of the tracking zero function Every TR T setting sampling perio...

Page 8: ...period 1 25 times sec 40ms 20 1 25 times sec 800ms 2 12 5 times sec 80ms 40 0 625 times sec 1 6s 4 6 25 times sec 160ms 80 0 3125 times sec 3 2s 8 3 125 times sec 320ms 100 0 25 times sec 4s 10 2 5 ti...

Reviews: