8. List of Paremeters
8.1 Condition data
Menu
Parameter
Default value
Protect
level
(*1)
Settable Selections / Ranges
Function / Remarks
P.L
Protect Level
PL0
PL0/PL1/PL2/PL3
Selects the protect level for preventing incorrect operation.
The higher the protect level, the more limitations are imposed
on a set parameter.
PVH
PH select
PH
PL0
PH/VH/PVH
Selects the type (peak hold, valley hold, or peak-valley hold)
that is activated when the PH function is enabled.
RANG
Input Range
14
PL1
11/12/13/14
Selects the input range.
25
22/23/24/25
AVG
Averaging times
1
PL1
1/2/4/8/10/20/40/80/100/200
Selects the number of averaging times (setting sampling rate).
By setting as the number of averaging times of inner sampling
of 25 times/sec (40 ms), the panelmeter practically uses the
average as input and acts with the period of “AVG × 40ms”.
Display of main display and output of BCD are also
synchronized with this setting sampling.
*See “10. Timing chart” for relationship the averaging times
and setting sampling.
MAV
Moving averaging
times
OFF
PL0
OFF/2/4/8/16/32
Selects the number of moving averaging times. (Lower filtering
effect OFF⇔2⇔4⇔8⇔16⇔32 Higher filtering effect)
S.WD
Step wide
1
PL0
1/2/5/0
Selects the resolution of the least significant digit. (When it is
set to “5”, the least significant digit indicates only “0” or “5”.)
BLNK
Display blank
level
OFF
PL0
OFF/B-3/B-2/B-1/ON
Selects display brightness. (Bright OFF⇔b-3⇔b-2⇔b-1⇔ON
Extinguished)
DL.T
Digital limitter
type
CUT
PL0
CUT/OVER
Selects display in case of overrange. When CUT is selected, the
set value of DLHI/DLLO is displayed; when OVER is selected,
O.L/-O.L is displayed.
BCD.L
BCD output logic
N.LOG
PL1
N.LOG/P.LOG
Selects the BCD output logic (N: negative logic, P: positive
logic).
* Only when BCD outputs are provided.
B. UP
DZ backup
OFF
PL0
OFF/ON
Selects whether to backup the digital zero value when power is
dicconnected.
LINE
Linearize
CLR
PL0
CLR/OFF/ON
Selects the enable (ON) /disable (OFF) of the linearize function
and data clear (CLR).
TR.T
Tracking zero
correction time
00
PL0
00~99
Sets the correction time of the tracking zero function. Every
“TR.T× setting sampling period”, the correction will be done.
In case of TR.T is “00”, the tracking function is disabled.
TR.W
Tracking zero
correction
width
01
PL0
01~99
Sets the correction width of the tracking zero function.
* Not available when TR.T is 00.
PON
Power on delay
OFF
PL0
OFF,1~30
Sets the time (set point x 1 sec.) taken from when the power is
turned on to the instant when measurement is actually started.
8.2 Scaling data
Menu
Parameter
Default
value
Protect level
Settable Selections / Ranges
Function / Remarks
FSC
Full-scale
display value
19999
PL1
-19999 ~ 19999
Sets the relationship between an input signal and display value.
FIN
Full-scale input
value
19999
PL1
-19999 ~ 19999
OFS
Offset display
value
0
PL1
-19999 ~ 19999
*See, ”Example of setting scaling data” in the section”4.7
Setting Scaling Data”.
OIN
Offset input
value
0
PL1
-19999 ~ 19999
DLHI
Digital limiter
High
19999
PL1
-19999 ~ 19999
Sets higher limit of displayable range. For higher input over
this setting, indications are not refreshed and kept the setting
value.
DLLO
Digital limiter
Low
-19999
PL1
-19999 ~ 19999
Sets lower limit of displayable range. For lower input below this
setting, indications are not refreshed and kept the setting
value.
DEP
Decimal point
.....
PL1
( Arbitrarily settable at each
digit)
Sets the display position of decimal-point.
If “.....” is set, no decimal-point is displayed.
(*1) When s
etting value of “P.L” is “0”, all parameters of PL0 to PL3 can be displayed and set.
When setting value of “P.L” is “1”, parameters of PL0 can not be displayed and parameters of PL1 to PL3 can be displayed and set.
When setting value of “P.L” is “2”, parameters of PL0 or PL1 can not be displayed and parameters of PL2 or PL3 can be displayed and set.
When setting valu
e of “P.L” is “3”, parameters of PL0 to PL2 can not be displayed and parameters of PL3 can be displayed and set.
(*2) Because all parameters in the scaling data is PL1,
when setting value of “P.L” is 2 or 3, scaling data setting mode can not be entered.
(i.e. “NET” is never displayed when “P.L” is set to 2 or 3.)