background image

Instruction Manual for MODEL A9

11-0

□,

A9

12-0

 

6/8 

5.3

 

Tracking Zero   

The  tracking  zero  is  a  function  for  automatically  digitally  correcting  the 
movement  of  the  zero  point  inside.  This  function  starts  to  work  at  the 
instant the digital zero function is enabled. Correction is made according 
to  the  values  set  for  the  tracking  zero  time  setting  and  tracking  zero 
width setting in the condition data.   

 

         

  (1) Digital zero function is enabled. Displayed value becomes 0. 

            (2)(3) At 5th sampling time, because displayed value is under 3 digit, 

  correction is performed and displayed value 

become “0”.   

            (4) Because displayed value is over 3 digit, correction is not 
                  performed. 

(5) Because displayed value is under 3 digit, correction is performed 

and displayed val

ue become “0”.

 

6. External Control Function   

For  those  equipped  with BCD  outputs,  there  are the hold,  digital  zero 

and peak hold functions that can be external controlled.   
The external control terminals are DC isolated from the power and input 
terminals.   

6.1 Hold Function   

The  hold function is  a function for stopping  refreshing  indication  at  an 

arbitrary timing. It is activated by shorting the HOLD terminal to the COM 
terminal or bringing its potential to the “0” level. 
 

6.2 Digital Zero Function   

The digital zero function is a function for resetting indication to zero at 

an  arbitrary  timing  and  then  displaying  a  range  of  variations  from  that 
point  onward.  ON/OFF  of  the  digital  zero  function  can  be  controlled 
either by the terminal control or by using keys on the front panel.   
For terminal control, this function is activated by shorting the DZ terminal 
to the COM terminal or bringing its potential to 

the “0” level.   

For  control  using  front-panel  keys,  it  can  be  activated  by  pressing  the 
Increment key with the Mode key held down. Taking the same step again 
causes this function to be deactivated. 
*  For  operation  using  the  control  terminals  or  the  front  panel  keys, 
terminal control has precedence over front-panel key operation. 

 
6.3 Peak Hold Function   

The  peak  hold  function  is  a  function  for  holding  the  maximum  value 
(Peak  Hold),  the  minimum  value  (Valley  Hold),  and  the  difference 
between  them  (Peak  Valley  Hold).  Switching  between  these  holding 
functions  is  achieved  using  condition  data.  The  peak  hold  function  is 
activated by shorting the P/H terminal to the COM terminal or bringing its 
potential to the “0” level. 

 
6.4 Pattern Select Function 

Pattern select function is a function selects one scaling data pattern from 
pattern 1to 4. 
A pattern is selected by the conditions of P.SEL0 and P.SEL1 terminals 
as shown below: 

Selected Pattern 

P.SEL1 

P.SEL0 

Pattern No.1 

Open / “1” level 

Open / “1” level

 

Pattern No.2 

Open / “1” level 

Short with COM / “0” level

 

Pattern No.3 

Short with COM / “0” level 

Open / “1” level

 

Pattern No.4 

Short with COM / “0” level 

Short with COM / “0” level 

* Only one pattern is required, leave P.SEL0 and P.SEL1 open and use pattern No.1. 

 

6.5 Control Terminal Signal Level 

Signal levels of all control terminals are as shown below : 

” level 

  0-1.5

V   

”1” level 

3.5-5

 

    (

Input current: -0.5

mA)

 

* Control terminals are isolated from the power input and signal inputs . 

7. Specifications 

  ■Input Specifications 
    ●DC voltage measurements                                                                  23℃±5℃,35 to 85% 

Range  Measurement 

range 

Display 

Accuracy 

Input 

impedance 

Maximum 

Permissible Input 

11 

±199.99mV  Offset 

±19999 

Full scale 

±19999 

±(0.1% of rdg + 2digit) 

100MΩ 

±50V 

12 

±1.9999V 

13 

±19.999V 

1MΩ 

±250V 

14 

±199.99V 

   

●DC current measurements                                                                    23℃±5℃,35 to 85%

 

Range  Measurement 

range 

Display 

Accuracy 

Input 

impedance 

Maximum 

Permissible Input 

22 

±1.9999mA  Offset 

±19999 

Full scale 

±19999 

±(0.2% of rdg + 3digit) 

10Ω 

±50mA 

23 

±19.999mA 

24 

±199.99mA 

0.1Ω 

±3A 

25 

±1999.9mA 

 

■Common specifications 

 

A/D  conversion   

:  ⊿Σ conversion method

 

Input circuit 

: Single-ended

 

Setting sampling rate 

: 25times/sec(max)   

 

Overrange warning 

: For an input signal exceeding the display range, displays 

  O.L.or –O.L..

 

Main display 

: 7-segment LED  (color: red, character height:14.2mm)

 

Display range 

-19999  ~  19999 

Zero indication 

: Leading zero suppression

 

Inner EEPROM 

endurance 

: 1,000,000 cycles              *1 

Operating 

temperature and 

humidity ranges 

: 0 to 50℃, 35 to 85 %RH (no condensation) 

 

Storage temperature 

and humidity ranges 

: -10 to 70℃, 60%RH or less

 

External dimensions 

: 96mm(W)  ×  48mm(H)  ×  75mm(D)

 

Weight 

: 160g (TYP) (AC power) / 150g (TYP)  (DC power) 

 

Dielectric strength 

: AC1500V for 1minute between the power terminals and each 

of  the  input,  BCD  outputs  and  the  external  control  (AC 

power). 

DC500V for 1minute between the power terminals and each 

of  the  input,  BCD  outputs  and  the  external  control  (DC 

power). 

DC500V  for  1minute  between  the  input  and  each  of  BCD 

outputs and the external control. 

AC1500V for 1minute between the casing and each terminal..

 

Insulation resistance 

:  100MΩ  or  more  at  500VDC  between  the  above-noted 

terminals,

 

  (*1)A writing to the internal EEPROM is performed, when the parameter setting is done and 

when the DZ(digital zero)input turns from OFF to ON if the digital zero backup is enabled. 
Please note that the number of the writing exceeds the endurance. 

 

■Power specifications   

 

●AC power  (A9111-0□, A9112-0□) 

Voltage range 

:  AC100 to 240V  ±10% 

 

Power consumption 

:  4.5VA (MAX)

 

●DC power  (A9311-0□, A9312-0□) 

Voltage range 

:  DC5V -5% to 12V +10%   

Power consumption 

:  1.5W (MAX)

 

●DC power  (A9411-0□, A9412-0□) 

Voltage range 

:  DC12 to 24V  ±10% 

 

Power consumption 

:  1.7W (MAX)

 

 

 

External control   

 

Hold 

:  Activated  by  shorting  the  HOLD  terminal  to  the  COM 

terminal or bringing the potential of the HOLD terminal to ”0” 

level. 

Digital Zero 

:  Activated by shorting the DZ terminal to the COM terminal 

or bringing the potential of the DZ terminal to ”0” level.

 

Peak hold 

:  Activated by shorting the PH terminal to the COM terminal 

or bringing the potential of the PH terminal to ”0” level.

 

Pattern select 

:  One of scaling data patterns is selected by a combination 

of  shorting  or  opening  P.SEL0and  P.SEL1terminals  to  the 

COM terminal   

* ”0” level : 0 to 1.5V with respect to COM terminal. ”1” level :3.5 to 5V with respect 

to COM terminal. 

 

  ■Option Specifications   

 

●BCD Outputs 

  ◎TTL output(A9□11-02, A9□12-02) 

Measured data 

:  Tri-state parallel BCD 

Polarity signal 

:  1 level for negative indication 

Overrange signal 

:  1 level for overrange indication 

Printout 

command 

signal 

:  Positive pulse output after the completion of measurement 

Output logic 

:  Switchable (PC logic not switchable) 

Output signal 

:  TTL level fan-out = 2, CMOS compatible 

  ◎Open collector output (NPN type)  (A9□11-03, A9□12-03) 

Measured data 

:    Negative logic.  (Transistor ON when is logic 1) 

Polarity signal 

:    Transistor ON for negative indication 

Overrange signal 

:    Transistor ON for overrange indication 

Printout 

command 

signal 

:    Transistor ON after the completion of measurement 

Output logic 

:  Switchable (PC logic not switchable) 

Transistor 

output 

capacity 

:  Voltage    30  V  DC  max.,  Current    10  mA  max.    Output 

saturation voltage    1.2 V or less at 10 mA   

   

●Enable   

function 

:  By  connecting  the  ENABLE  terminal  to  COM terminal  or 

bringing  the  potential  of  the  ENABLE  terminal  to”0”  level, 

BCD  outputs  become  High  impedance  (TTL  output)  / 

transistors turn off (Open collector output)

 

*  “0”  level  :  0  to  1.5V  with  respect  to  COM  terminal,“1”  level  :  3.5  to  5V  with 

respect to COM terminal   

 

Tracking 
zero 
starts 

5 samplings

 

(Tracking 

zero time)

 

Digital zero ON

 

Displayed 
value

 

(1)

 

(2)

 

(3)

 

(4)

 

(5)

 

0

 

-1

 

-2

 

-3

 

1

 

2

 

3

 

(Display) 

 

Example of setting:  Tracking zero time setting 5 (Correction made every 5 averaging (sampling))

 

 

        Tracking zero width setting    3(correction width of 3digit)

 

Tracking zero 
width

 

Summary of Contents for A9000 Series

Page 1: ...12 22 or 23 range Input terminal HI input terminal of 13 or 14 24 or 25 range Input terminal LO input terminal Selecting input range can not be done only by connecting input signal to a suitable termi...

Page 2: ...vel 3 5 5V Input current 0 5mA 4 Parameter Settings 4 1 Components and Functions Before setting parameters remove the front panel by inserting a flat blade screwdriver into the ditch under the front p...

Page 3: ...item 5 Press the Mode key to move to peak hold type setting 6 Set up the peak hold type using the Increment key This setting example sets peak hold type to PH PH Holds the maximum display value PV Ho...

Page 4: ...Enter and Mode keys during measurement operation to enter the condition data setting mode 2 Press the Shift key to move to the scaling data setting mode 3 Press the Mode key to display the number of...

Page 5: ...the setting again The number of linearize points is up to 16 but a value 17 to 19 is also displayed during setup Note that if you set a value from 17 to 19 it is forced to set to 16 5 Other Functions...

Page 6: ...23 19 999mA 24 199 99mA 0 1 3A 25 1999 9mA Common specifications A D conversion conversion method Input circuit Single ended Setting sampling rate 25times sec max Overrange warning For an input signa...

Page 7: ...ble ON disable OFF of the linearize function and data clear CLR TR T Tracking zero correction time 00 PL0 00 99 Sets the correction time of the tracking zero function Every TR T setting sampling perio...

Page 8: ...period 1 25 times sec 40ms 20 1 25 times sec 800ms 2 12 5 times sec 80ms 40 0 625 times sec 1 6s 4 6 25 times sec 160ms 80 0 3125 times sec 3 2s 8 3 125 times sec 320ms 100 0 25 times sec 4s 10 2 5 ti...

Reviews: