Process documentation
101
Communication module for external archiving
A module slot is integrated into the back of the machine for a W&H
communication module to permanently archive cycle reports. The
module enables the installation of an Ethernet interface for docu-
mentation using documentation software or an RS-232 interface for
connection to a report printer.
Please contact W&H for further information on software and suitable
printers.
Only use terminal devices (PC, printers etc.) which comply with EN/
IEC 60950.
The communication modules are available from W&H as an access-
ory and can be retrofitted at any time. The modules are supplied with
their own installation instructions.
The interface must only be configured by suitably qualified and com-
petent persons. Follow the instructions in “Further settings/Interface”.
Process documentation using external software
For digital archiving the process data is transmitted to external docu-
mentation software via an Ethernet interface. Transmission can op-
tionally occur continuously during the process or as a single transac-
tion at the end of the process. The settings for this are modified by
W&H Service.
Information on wash pressure, A
0
value, conductivity and temperature
in the wash cabinet can be archived graphically if required.
Installation of an Ethernet interface requires the retrospective fitting of
an XKM 3000 L Med communication module.
For connection to a WLAN network the module can be connected via
a cable to an existing wireless access point.
Problems with
data transmission
If there is a network problem during a running process, e.g. due to a
loose cable, a relevant fault message is displayed.
Network down
The process running will be continued without interruption and the
process data will be saved in the meantime in the internal memory.
In the event of network or report software problems contact your sys-
tem or network administrator.