Maintenance
Lexa PLUS | Instructions for Use | RIS-311 ENG Rev03 | 29/08/2022 | © W&H Sterilization Srl
69
400-cycle maintenance
REPLACING THE HEPA FILTER
Notice
: The HEPA filter needs to be replaced every 400 sterilization cycles or once a year, whichever come first. A replacement message alerts
when replacement is due. If the consumables is replaced prior to the message, you have to reset the consumable cycle counter.
Proceed as follows:
1
Open the sterilizer door.
2
Unscrew the HEPA filter by hand (counter-clockwise).
3
Screw on the new HEPA filter (clockwise) and tighten it snug.