Diagnostics
Lexa PLUS | Instructions for Use | RIS-311 ENG Rev03 | 29/08/2022 | © W&H Sterilization Srl
83
RESET THE SAFETY THERMOSTAT
The sterilizer is fitted with a safety thermostat to prevent it from overheating. If the safety thermostat operates because of too high temperatures,
the error 240 or a timeout error is displayed. The thermostat must be reset manually. Proceed as follows:
1
Switch the sterilizer OFF and remove the mains cable.
2
Wait for the sterilizer to cool down.
3
Open the sterilizer door.
4
Unscrew the cap and push on the reset button of the thermostat switch; a click
sound indicates that the thermostat switch has been reset.
5
Screw the cap.
6
Connect the mains cable and switch the sterilizer ON.
Note
: If the thermostat operates repeatedly, call technical service.
7
Wait for the sterilizer to finish the error reset phase and follow the instructions on
the display.