Maintenance
70
Lexa PLUS | Instructions for Use | RIS-311 ENG Rev03 | 29/08/2022 | © W&H Sterilization Srl
800-cycle or biannual maintenance
SEQUENCE OF PROCEDURES TO CLEAN THE WATER TANK
To clean the water tank, proceed as follows:
1
2
"Access to the water tank" on the next page" "Access to the water tank" on the next page.
3
"Clean the water tank" on page 72
PREPARE THE STERILIZER FOR CLEANING THE WATER TANK
Proceed as follows:
1
Switch OFF the sterilizer and disconnect the mains cable.
2
Completely drain both water tanks (see "Draining the used and clean water tank"
on page 75).
3
To drain the detergent solution during the subsequent cleaning, leave the drain
tube attached to the drain connection of the tank you want to clean.