![W&H Lexa PLUS Instructions For Use Manual Download Page 72](http://html1.mh-extra.com/html/wandh/lexa-plus/lexa-plus_instructions-for-use-manual_3076914072.webp)
Maintenance
72
Lexa PLUS | Instructions for Use | RIS-311 ENG Rev03 | 29/08/2022 | © W&H Sterilization Srl
CLEAN THE WATER TANK
Notice
: Do not touch the water level sensors. If misplaced or misaligned from their original position, the operation of the sterilizer could be
impaired.
Proceed as follows:
1
Clean the internal tank surfaces with a soft sponge moistened with water. For
better cleaning results, clean with W&H MC-1000 cleaning solution. Use a small
non-abrasive brush to clean the areas that are difficult to reach.
Notice
: never use any other disinfectant, detergent or abrasive product, as they
might result aggressive for the tank material.
2
Rinse the internal tank surfaces thoroughly, until all residuals of dirt and
detergent have been eliminated.
3
Dry the internal tank surfaces.
5
Remove the internal filter.
6
Clean the metal cartridge of the internal filter with tap water.
7
Reposition the internal filter.
8
Detach the drain tube.