Lexa PLUS | Instructions for Use | RIS-311 ENG Rev03 | 29/08/2022 | © W&H Sterilization Srl
65
Maintenance
Frequency
1
Cycles
1
Operation
Yearly
2
400
2
Replace the HEPA filter. See "400-cycle
maintenance" on page 69.
Yearly
2
1200
2
Replace the door gasket. See "Replacing the door
gasket" on page 73
1
: whichever occurs first.
2
: even if the maximum cycle number is not reached, it is recommended to
replace the consumable parts every year, or if they appear worn or
damaged, or if the filters are clogged or discolored.
EXPIRED MAINTENANCE
The sterilizer monitors the wear of consumables by counting the
number of cycles executed since the last replacement.
When the number of cycles is close to the maximum, a pre-alert
about the concerned consumable is displayed. Please check that
you have the requested spare part available, buy one if not. When
the maximum number of cycles has been met, a message to replace
the consumable will be displayed.
If you can not replace the consumable immediately, the sterilizer
will operate anyway but the message will appear again some cycles
later.
1
Tap to see an animated replacement procedure.
2
When you have replaced the consumable tap to confirm: the
executed cycle counter is reset.
REPLACE THE CONSUMABLE BEFORE THE MAINTENANCE DUE
DATE
If you replace the consumables before the request of replacement
appears, you should manually reset the counters through the
following procedure.
1
On the homepage, tap
>
2
Select the consumable you want to replace: a message appears
showing the current worked hours of the part.
3
Tap
to see an animated replacement procedure.
4
When you have replaced the consumable tap
to confirm: the
executed cycle counter is reset.