4.0 SETTING THE OPERATING PANEL & USE
QE SEPCOM SIP BED EASY
4.5.1 HOW TO RESET THE WARNING ALARMS
4.5.1.1 Maintenance performed
In case of maintenance, follow the procedure bellow to reset the alarm:
1.
Log in to the PLC of the control panel with ADMIN credentials (technician)
2.
Tap ON TECHNICIAN SETTINGS
3.
Use the right side arrow of the menu to go the page that displays the operation hours of the users.
4.
Reset manually the operation hours (left column) using the numeric keypad (check the section
How to
reset the operation hours
).
The red flag on the screen edge will automatically disappear; now you can start a new work cycle without
interruption.
4.5.1.2 Maintenance not performed
If maintenance cannot be performed, the alarm can be by-passed and a new cycle can be started. At the end
of this cycle the alarm will trip again. Follow the steps indicated below:
1.
Press the STOP cycle button
2.
Press START cycle button to restart.
Danger - Warning
The next time the alarm trips, the system will be locked again.