Operating Instructions | symphony 5.3 A – 06/2009
61
12
Technical data
Mechanical / thermal data:
Description
Unit
Value
External dimensions
Width x Depth x Height
mm
in
637 x 867 x 766
25.1 x 34.1 x 30.2
Interior dimensions
Width x Depth x Height
mm
in
470 x 607 x 530
18.5 x 23.9 x 20.9
l
approx. 151
cu.ft
approx. 5.3
Shelves
Width x Depth
mm
in
423 x 455
16.7 x 17.9
Standard quantity
Piece
3
Maximum quantity
Piece
10
Maximum surface load
Piece
10 / insertion shelf
Maximum device overall load
kg
ib
30
66.1
Weight, without accessories
kg
ib
70 (stainless steel)
154.3 (stainless steel)
Ambient temperature range
°C
+18...33
Temperature control range
°C
RT + 3 ... 55
Temperature deviation at 37°C, time
°C
± 0.1
Temperature deviation, spatial (DIN 12880, Part 2)
at 37 °C
°C
± 0.6
Duration of the auto-start routine, to 37 °C
ambient temperature 20 °C
h
5 ... 10
Temperature recovery time, at 37 °C, door open 30
seconds (to 98 % of initial value)
min
< 10
Heat transfer to environment:
at 37 °C
during 90 °C decontamination
kWh/h
kWh/h
0.06
0.26
Water quality
demineralized / distilled or
autoclaved or completely
deionized / distilled or
autoclaved.
Liquid quantity:
Incubation operation
90°C decontamination operation
l
ml
max. 3.0 / min 1.2
300
Constant humidity at 37 °C
% rH
approx. 93
Humidity recovery time, at 93 % rH, door open 30 s
(to 95 % of initial value)
min
min
approx. 20 (stainless steel)
Thermal
Humidity
Mechanical
Chamber volume