Installation of the device
Operating Instructions | symphony 5.3 A – 06/2009
3
Installation of the device
3
Installation of the device
16
3.3 Space requirements
A
Fig.1:
When installing the device, make sure that the
installation and supply connections are freely accessi-
ble.
The control box at the rear panel of the device may
serve as a spacer to adjacent objects. The side
distances given are minimal distances.
To protect the CO
2
incubator against contamination,
use a floor stand even if the device is installed near the
floor. The height of the floor stand should not be less
than 200 mm.
Several floor stands and carriers are available as opti-
ons from VWR (part numbers see Section 11.1, “Parts
and accessories”).
NOTE!
Accessibility of the devices:
To ensure the accessibility for care and
maintenance works, keep larger side and
rear distances!
3.4 Transport
Fig.2:
For transport do not lift the device using the
doors or components attached to the device (e.g. con-
trol box on rear panel) as lift points.
NOTE!
Lift points:
Lift the device only using the lift points
shown in Fig. 2!
x
75
100
100
669
870
637
766
F
F
F
³
185
Fig.1: Device dimensions
Fig.2: Lift point