59
vwr.com I Instruction manual VWR® U Water System
d)
Tilt the upper end of the UV lamp chamber out.
e)
Wear gloves included in the new UV lamp package. Avoid direct skin contact with the quartz
glass of the UV lamp. Remove the UV lamp cover to expose the UV lamp. Unplug the UV
lamp from its power cord, and then carefully remove the old UV lamp from the chamber.
f)
Carefully insert the new UV lamp into its chamber. When about 2/3 of the lamp is inserted,
hold the UV lamp and connect it to the ballast cable connector (4-pin connector) as shown in
the picture, and then gently insert the new UV lamp completely into the chamber.
g)
Cap the UV chamber with the black mask.
h)
Install the UV lamp chamber back to the original position.
i)
Turn on the system power. Scan the RFID chip for the UV lamp on the system. Tap
Install
on
the pop-up dialog
to reset the corresponding UV lamp.
5.4.3 Replacing the Tank UV Lamp