3.2.6 Install Consumables
3.2.6.1 Register Consumables (Not including cartridges)
Follow instructions on the screen, scan RFID tags
for the RO Pack, T Pack (optional), 185 nm UV
lamp, tank UV lamp, tank vent filter and final filter
at the AC pack position with an antenna sign
( ) on the system. Details are shown below
Note:
1.
RFID tags are either inside consumable
packages, or for preinstalled UV lamps etc,
inside USB key package.
2.
For traceability, your User Name is required to
install/register consumables. For a new
installation, enter system manager’s User
Name.
Remove the right side panel to find the RFID
reader position ( ) as shown left.
A pop-updialog will be shown as left after every
scan, enter manager’s User Name and tap
Install
to register the consumables
Follow steps below to install cartridges after
finishing the scan.