Display
The display provide some additional features:
·
LCD display with 3 rows of real time data, refreshed every second
·
Key pad with menu to configure the main settings
·
Data logger with adjustable logging intervals (option)
·
5 custom units (multiply and existing unit with a factor and show it on the LCD display)
7.1
Display status icons
Some status icons show feedback on the meters' status. Below is a list with explanation
Icons
Description
Sensor module is properly connected and supplied with power
No communication with the sensor [Check external power when disconnected]
A blinking dot will indicate that a data session is active
2 rotating arrows indicate that there is communication with the computer
The display is locked. The menu can not be accessed
Memory indication. Each block indicates 20% of memory usage. The blocks start to
blink if the memory is more then 95% full
7.2
LCD display
The LCD display provides 3 rows for real time data. Each row can be configured in the menu by
selecting the desired parameter for this row.
Available options are listed in
menu -> display
.
7.3
Data Logger
The optional integrated data logger offers you 2 Million data points. Enough to measure all three
channels 1 x per second for more than a week.
Use the following guidelines for the intervals
Application
Flow
Pressure
Temperature
Standard energy management application
5 min
5 min
5 min
Machine testing - quick fluctuations
1 sec
1 sec
1 sec
Audit - one week
10 sec
10 sec
5 min
Audit - one month
30 sec
30 sec
5 min