© 2017 Van Putten Instruments BV | MAN-VP-SINL-UK | Revision:1701 | Date:22-03-2017
9
4
Quick start
This chapter contains the basic steps to start using your VPFlowScope in-line flow meter. Additional
information on all subjects can be found in the next chapters.
1. Unpack
Unpack the meter and check if all items are there and in good shape. A checklist with all items is
available on the box.
2. Mechanical installation
·
Find the best point of installation for this product. Make sure that all specifications are met.
·
For installation of the VPFlowScope in-line, the pipe needs to be cut. Mount the VPFlowScope in-
line between the pipe ends. Use tri-clamp or similar adapters for quick installation and removal.
·
For installation with the tubing kit: The connection is BSP outer straight thread [For NPT thread,
chase the BSP thread of the pipe ends with a die]; 0.5", 1" or 2" depending on the VPFlowScope
in-line model.
·
For installation without tubing kit: The connection is BSP inner straight thread [It can accept NPT
male thread ends]; 0.5", 1" or 2" depending on the VPFlowScope in-line model.
See chapter
mechanical installation
for more detailed information.
3. Electrical installation
3.1 Permanent installation
Connect a cable with 5 Pin M12 connector to the VPFlowScope in-line. The cable can be
connected to a central data acquisition / building management system or data logger via
Modbus, 4..20mA or pulse. See chapter
electrical connections
for more information.
Apply 12-24 VDC to power up the device. Use a Class II power supply (less than 2 Amps). If the
built-in display option is available, it will light up when power is applied.
3.2 Temporary installation
Use a 12 Volt power supply with M12 connector to power the VPFlowScope in-line. This quick
method is ideal for audits.
4. Data recording
When the data logger is available a data log session can be started by pressing the esc button and
then enter. All parameters will be logged with the default logging intervals. These logging intervals can
be changed with the VPStudio software. This software tool is also used to retrieve the recorded
sessions.