32
• Connect the two measuring prods with the object to be measured
The measuring value is shown on the analogue scale “1”. Take the value and multiply it with
the values of the following table:
dC voltage range
Analogue reading scale
Multiplication factor
2 V
2
x 1
10 V
10
x 1
50 V
50
x 1
100 V
10
x 10
250 V
250
x 1
500 V
50
x 10
100 mV
10
x 10
Alternating Current Measurement
• Connect the black measuring line to the COM socket “8” and the red one to the socket V, Ω,
“7”.
• Set the adjustment wheel “4” to the area for voltage measurement and the slider “5” to “AC“
• Connect the two measuring prods with the object to be measured.
• The measuring value is shown on the analogue scale “1”. Take the value and multiply it with
the values of the following table:
AC voltage range
Analogue reading scale
Multiplication factor
10 V
10
x 1
50 V
50
x 1
250 V
250
x 1
500 V
50
x 10