7
Power input connector
T
ECHNICAL
S
PECIFICATIONS
Flow detection principle
Fleisch type pneumotachometer
Back pressure
0.1kPa/litre/second, complies with or exceeds
ATS standards.
Volume measurement
Flow integration
Maximum recorded flow rate
±
14l/s
Maximum recorded volume
>8 litres
Measuring accuracy
Better than
±
3% or
±
0.05l volume whichever is
greater and
±
5% flow (inspiratory:
±
5% vol.,
±
8%
flow)
Operating temperature
15-40ºC
Safety standards
Designed to conform to EN60601-1, EN60601-1-2
Spirometric standards
ATS and ERS
Gross Size
170mm x 163mm x 45mm
Gross Weight (without
flowhead)
345g
Storage Temperature
0-50ºC
Storage Relative Humidity
10% - 95%
CE Notice
Marking by the symbol
l
indicates compliance of the Vitalograph Pneumotrac
to the Medical Devices directive of the European Community. Such marking is
indicative that the Vitalograph Pneumotrac meets or exceeds the following
technical standards:
•
EN60601-1 - “General requirements for safety”
•
EN55011 - “Radiated and mains conducted emissions for industrial, scientific
and medical (ISM) radio-frequency equipment”
•
IEC801-2 - “Electromagnetic compatibility for industrial-process measurement
and control equipment Part 2: Electrostatic discharge requirements.”
•
IEC801-3 - “Electromagnetic compatibility for industrial-process measurement
and control equipment Part 3: Radiated electromagnetic field requirements.”
FDA Notice
Caution: Federal Law restricts this device to sale by, or on the order of a
physician.