Technical information
Technical information
Failure en-
countered
Possible cause
Corrective action
Irregular heartbeat,
disruptive move-
ments, shaking,
wobbling (objects),
breathing in deeply,
etc.
Repeat measurement
after 3-5 minutes rest.
Possible factors (depend-
ing on the severity of the
arrhythmia) affecting the
measurement results
must be discussed with
your doctor.
Measured
values /
pulse values
are extreme-
ly high or
low
Too little rest before
the measurement.
Movement or talking
during measure-
ment.
Repeat measurement af-
ter a break of approx. 3-5
mins. Observe instruc-
tions for use Page 43.
Incorrect measure-
ment position.
Repeat measurement,
ensuring correct body
posture. (Page 43)
Consecutive
measure-
ments
produce
different
measured
values (read-
ings).
Blood pressure is
not a fixed quantity.
It may vary in the up-
ward or downward
direction within a
few minutes by more
than 20 mmHg.
Always take the measure-
ments in the same con-
ditions in order to obtain
comparable results.
Measure-
ment was
interrupted
and cuff first
deflated and
then inflated
Unit detects a fault
or that diastolic
value is too low.
Unit shows no fault. Re-
peat measurement.
Movement during
measurement
If necessary, interrupt
measurement, measure
again after a break of 5
minutes.
Failure en-
countered
Possible cause
Corrective action
No display
after unit is
switched on
Batteries inserted
incorrectly.
Check position of bat-
teries.
Batteries flat.
Change batteries.
Battery contacts
dirty.
Clean battery contacts.
No display
after Memo-
ry button is
pressed
No measured values
stored.
Take a new measure-
ment.
Batteries flat
Change batteries.
Measure-
ment inter-
rupted
Batteries flat
Change batteries.
Date/time
not shown
in display
- Function accidental-
ly deactivated
- Function was not
reactivated after
battery replacement
Activate date and time as
described on page 45.
2. Customer service
The unit may only be repaired by the manufacturer or by an
expressly authorised organisation. Please contact: