Commissioning
6.3.6 Connect the digital I/O interface (Digital I/O card ADDI-
DATA)
NOTICE
Damage of the outputs through inductive loads
•
With induced loads the outputs have to be switched with a free-wheeling
diode anti-parallel to the load.
NOTICE
Destruction caused by potential differences
•
The ground potential (GND) of the digital inputs and outputs of the device
must be connected to the ground potential (GND) of the control voltage of
the device to be controlled.
Connecting
Die zum Anschluss benötigten Komponenten erhalten Sie als Zubehör.
"Accessories for the option Digital I/O card ADDI-DATA", page 12
Connect the 37-pin D-Sub connector of the device's I/O expansion card to the
process environment using one of the following two variants:
•
I/O cable with 2 x Sub-D and terminal board
•
I/O cable with 1 x Sub-D and open cable end