Technical Manual
-4
General errors
These errors generally occur due to faulty entries from the front panel, alternatively
invalid data or not allowed commands from the control unit.
Error
code
Explanation
008
Input channel not used
The Input channel is not configured (used) in the instrument.
009
Function Block not used
The Function Block is not configured (used) in the instrument.
060 -
063
AOUT ch X termination error
There is a termination error on an analog output channel. Too low
impedance on a voltage output or too high impedance on a current output
may cause this error.
070
I/O module communication error
There is an internal error in the communication between the CPU module
and one or more of the I/O modules. Restart the instrument and contact the
supplier if the error persists.
071 -
074
I/O module reset
There has been an internal reset within an I/O module. Restart the
instrument and contact the supplier if the error persists.
100
Instrument in wrong state
The transmitted command is not applicable to the present instrument mode.
101
Too high value!
Value over allowed range. Compare with restrictions for the parameter.
102
Too low value!
Value under allowed range. Compare with restrictions for the parameter.
103
Illegal start address.
Illegal Modbus start address, when writing data to the G4 instrument.
104
Illegal number of registers.
Illegal number of Modbus registers, when writing data to G4 instrument.
105
Faulty value
Value not accepted when entering a parameter from front panel or illegal
data in Modbus registers, when writing data to the G4 instrument.
113
Outside zero setting limits.
Any Force value in a Function Block is outside allowed zero setting range.
118
Zero setting not allowed.
Zero setting is not allowed when the calibration menu is displayed for the
actual Function Block.
240
Fieldbus command error.
Command error response on fieldbus connection. Indicates that the
command sent to the G4 instrument via the fieldbus could not be executed.
249
Internal Timeout
A internal timeout occurred. If this error occurs when a command is sent to
the instrument, try the command again. Restart the instrument or contact the
supplier if the error persists.
Summary of Contents for G4-DT
Page 2: ......
Page 18: ...Technical Manual 1 14 ...
Page 36: ...Technical Manual 2 18 ...
Page 45: ...G4 Multi Channel Force Instrument 4 5 ...
Page 76: ...Technical Manual 4 36 ...
Page 82: ...Technical Manual 5 6 ...
Page 92: ...Technical Manual 6 10 ...
Page 138: ...Technical Manual 7 46 ...
Page 158: ...Technical Manual 8 20 ...
Page 164: ...Technical Manual 9 6 ...
Page 170: ......
Page 171: ...G4 Multi Channel Force Instrument Appendix 1 ...
Page 172: ...Technical Manual ...
Page 173: ......