Technical Manual
7-24
Data representation
Data sent to and from the instrument uses 16 bit holding registers (40XXX) and can
use different formats for flexibility.
Integer
Unsigned integer (1 modbus register)
Values stored in one modbus register as an unsigned integer
(16 bit number without decimals).
Scaled integer (2 modbus reg 1 modbus register = 3 modbus registers)
Values stored in a special 3 register format. The first two registers are used as
a 32 bit long integer value (with sign) and the third register is holding the number of
decimals in the value.
Example: 12345678 (32 bit number) in the two first registers and 3 in the third register
gives the value: 12345.678.
Register
Hex
Decimal Description
1
00BC
188
The 16 most significant bits in the value.
2
614E
24910
The 16 least significant bits in the value.
3
0003
3
The number of decimals.
Calculations in decimal numbers:
First multiply the most significant register with 2
16
(65536) and add
the least significant register to the value.
188 * 2
16
+ 24910 = 12345678
Now divide the number to get the right number of decimals. The decimal register was
set to 3 in this example, which gives the value 10
3
= 1000 to divide with.
12345678 / 1000 = 12345.678
Summary of Contents for G4-DT
Page 2: ......
Page 18: ...Technical Manual 1 14 ...
Page 36: ...Technical Manual 2 18 ...
Page 45: ...G4 Multi Channel Force Instrument 4 5 ...
Page 76: ...Technical Manual 4 36 ...
Page 82: ...Technical Manual 5 6 ...
Page 92: ...Technical Manual 6 10 ...
Page 138: ...Technical Manual 7 46 ...
Page 158: ...Technical Manual 8 20 ...
Page 164: ...Technical Manual 9 6 ...
Page 170: ......
Page 171: ...G4 Multi Channel Force Instrument Appendix 1 ...
Page 172: ...Technical Manual ...
Page 173: ......