Technical Manual
7-30
Bytes 2 and 3:
Define from which register number to read or write.
Bytes 4 to 15:
Contains the data when writing to the instrument.
Data to the fieldbus (Inputs in the master)
General
Data to the fieldbus are divided into two categories, one is the mandatory “base block”
which contain general data and the 12 modbus register used for reading from the
instrument. The second part is the up to twelve configurable data blocks.
Byte
Contents
Byte
Contents
00
Instrument error
MSB
16
Read register 5
MSB
01
Instrument error
17
Read register 5
02
Instrument Status
MSB
18
Read register 6
MSB
03
Instrument Status
19
Read register 6
04
Command acknowledge
20
Read register 7
MSB
05
Command error
21
Read register 7
06
Start address, Read
MSB
22
Read register 8
MSB
07
Start address, Read
23
Read register 8
08
Read register 1
MSB
24
Read register 9
MSB
09
Read register 1
25
Read register 9
10
Read register 2
MSB
26
Read register 10
MSB
11
Read register 2
27
Read register 10
12
Read register 3
MSB
28
Read register 11
MSB
13
Read register 3
29
Read register 11
14
Read register 4
MSB
30
Read register 12
MSB
15
Read register 4
31
Read register 12
Summary of Contents for G4-DT
Page 2: ......
Page 18: ...Technical Manual 1 14 ...
Page 36: ...Technical Manual 2 18 ...
Page 45: ...G4 Multi Channel Force Instrument 4 5 ...
Page 76: ...Technical Manual 4 36 ...
Page 82: ...Technical Manual 5 6 ...
Page 92: ...Technical Manual 6 10 ...
Page 138: ...Technical Manual 7 46 ...
Page 158: ...Technical Manual 8 20 ...
Page 164: ...Technical Manual 9 6 ...
Page 170: ......
Page 171: ...G4 Multi Channel Force Instrument Appendix 1 ...
Page 172: ...Technical Manual ...
Page 173: ......