7.4.4.8 Ignore or Accept Sync (Global) Limits
Length
Command
Group
Command
Code
Card Mask
Channel
Mask
Parameters
0x0007
0x04
0x0008
Used
Not Used
Byte 8:
0 – ignore sync (global) limits
and do not record (default)
1 – accept global limits
This command sets the card to either ignore or accept synchronized (global) limits. If global limits are
ignored no recording will be done if a sync (global) limit is detected
7.4.4.9 Set Pre-trigger Buffer Size Command for Sync (Global) Limits
Length
Command
Group
Command
Code
Card Mask
Channel
Mask
Parameters
0x000A
0x04
0x0009
Used
Ignored
0x00
Bytes 8-11:
32-bit buffer size, in scans
Sets the “pre-trigger” buffer size for synchronized (global) limits.
Range: 0 to 645,000 (or the size of the scan buffer)
Default: 0
7.4.4.10 Set Post-trigger Buffer Size Command for Sync (Global) Limits
Length
Command
Group
Command
Code
Card Mask
Channel
Mask
Parameters
0x000A
0x04
0x000A
Used
Ignored
0x00
Bytes 8-11:
32-bit buffer size, in scans
Sets the size of the “post-trigger” buffer for synchronized (global) limits.
Range: 0 to 4294967295