Repairing
Replace measuring head
Procedure
- Loosen clamps on the main connector plug
- Pull off the main connector plug
- Dismantle the measuring head (8 screws, see fig. 6. 01 / 1)
- Examine if the measuring head is equipped with the correct factory supplied
(or the one required for the system) wire break resistor values.
To change the wire break resistors, see instructions on the protecting cover on the
back side of the measuring head. (see fig. 6. 02 / 1)
More information can be found in chapter 2, installation, under
"electrical connection".
In case that the value of the wire break resistors do not match your installation :
- Change the wire break resistors of the new measuring head. The resistors are
designed as plug-in resistors and can be replaced without soldering !
- Fasten measuring head, tighten 8 screws moderately
- Push on main connector plug
- Lock into place
- Carry out performance test according chapter 4
Attention!
replacement of wire break resistors must only be done by skilled and trained
personnel!
6. 01
Operating Instruction Visatron VN 93
06. 04
6. 01 / 1
Key width 4mm
Summary of Contents for 10980
Page 2: ...06 04...
Page 20: ...2 05 operating Instruction Visatron VN 93 06 04 2 05 1 Installation of siphon blocks...
Page 22: ...Installation of siphon blocks 2 07 operating Instruction Visatron VN 93 2 07 1 06 04...
Page 23: ...Installation of siphon blocks 2 08 operating Instruction Visatron VN 93 2 08 1 06 04...
Page 68: ...Brochures Leaflets etc 10 01 Operating Instruction Visatron VN 93 06 04...
Page 70: ...Notes 12 01 Operation Instruction Visatron VN 93 06 04...
Page 71: ...Notes 12 02 Operating Instruction Visatron VN 93 06 04...
Page 72: ...Notes 12 03 Operating Instruction Visatron VN 93 06 04...
Page 73: ...Notes 12 04 Operating Instruction Visatron VN 93 06 04...