vinmetrica SC-100A User Manual Download Page 9

SC-100A manual 

Version 3.1c 

Yello

LED

Section 

Equipment Required 

Description 

 

6.SO

2

 

Voltage 

SO

2

 probe or  

SO

2

 probe simulator. 

The raw voltage output from the 
instrument's current amplifier is 
displayed as X.XX volts. (.XXX if less 
than 1.00) 

 

7.DAC Test 

None.   
Disconnect probe. 

The Digital-Analog Converter (DAC) is 
cycled through its 32 levels.   

 

8.Battery 
Voltage 

Install two AA 
batteries 

The battery voltage is displayed as X.XX 
volts. 

 

9.Character 
Set 

None. 

Every ASCII character (space) to ~  is 
displayed.  Due to the limitations of the 
7-segment format, some characters are 
not used by the software. 

 

10. Number 
Display 

None. 

The display cycles through showing 
every possible digit and every decimal 
point. 

 

11.Sound 
Test 

None. 

The beeper is turned on continuously. 

 

12. pH CAL 
values  

None. 

[only  firmware 

v3.0.6 and later]

 

[NA] 

Displays current CAL values for 

pH 7 and 4, in mV.  Pressing ENTER 
cycles between these.  If a pH CAL reset 
has been done, displays CAL value for 
pH 3 rather than 4. 

 

 

13.

 

pH CAL 

reset 
 

None.

  [only  firmware 

v3.0.6 and later]

 

[NA] 

Displays “

PrESS EntEr

”; Press 

ENTER to reset pH CAL parameters to 
default values.  Message “

Good CAL 

rSt

” then scrolls.

 

 

14.

 

 pH DAC 

Set 
 

None.

  [only  firmware 

v3.1.2 and later]

 

[NA] 

Displays “

PH DaC Set

” then 

displays the DAC index for pH bias 
voltage, followed by the voltage value 
itself.

 

 

15.

 

SO

2

 

Baseline 

None. Disconnect SO

2

 

electrode 

[only firmware 

v3.1.0 and later]

 

Sets baseline value for SO2 mode.  Let 
message scroll 5 sec., during which time 
of the baseline is automatically set.  
Normal values are 0.02 to 0.10 

 

16.

 

 TA 

endpoint 

None 

[only  firmware 

v3.1.2 and later]

 

[NA] 

Displays “

seT Ta PT

” 

then 

displays TA endpoint value. Press ENTER to 
toggle between default of 8.2 (USA standard) 
or 7.0 (European standard).

 

 

Summary of Contents for SC-100A

Page 1: ...ge 5 Two 3 mL polyethylene transfer pipettes 6 One 25 mL serological pipette 7 One 100 mL polypropylene beaker Things you will need 1 Two standard AA batteries alkaline type 2 Distilled water which us...

Page 2: ...63 79 99 125 pH 3 0 3 1 3 2 3 3 3 4 3 5 3 6 3 7 3 8 3 9 4 0 We recommend using a sulfite calculator for determining how much sulfite to add to your wines after taking a sulfite measurement with the SC...

Page 3: ...100A shuts off after 30 minutes If this happens unexpectedly just press the POWER button to resume from where you were 4 The SO2 electrode is attached via the BNC connector protruding from the top To...

Page 4: ...to dispense the SO2 Titrant into the top of the burette 3 Using the 25 mL pipette provided in the kit place 25 mL of wine or must in the titration beaker Figure 5 Draw sample up to the 0 mL mark then...

Page 5: ...lectrode Holder adjust it to a similar level You may add up to 20 mL water if needed 8 Verify that the current is less than 50 and the green PROCEED LED is lit If the current is greater than this and...

Page 6: ...gent and 2 mL of the Reactant solution to the vessel Remember that if you are using the transfer pipettes in the kit 2 mL is marked on the side of the neck of the transfer pipette so dispense four of...

Page 7: ...METRICA S MAXIMUM LIABILITY FOR ALL DIRECT DAMAGES INCLUDING WITHOUT LIMITATION CONTRACT DAMAGES AND DAMAGES FOR INJURIES TO PERSONS OR PROPERTY WHETHER ARISING FROM VINMETRICA S BREACH OF THESE TERMS...

Page 8: ...st Mode hold the POWER button down 5 10 seconds until the instrument shuts off If the device does not shut off after 10 seconds of holding down the button move to the next section by releasing then pr...

Page 9: ...2 pH CAL values None only firmware v3 0 6 and later NA Displays current CAL values for pH 7 and 4 in mV Pressing ENTER cycles between these If a pH CAL reset has been done displays CAL value for pH 3...

Page 10: ...r this type of wine molecular SO2 pH is your value mg L Redo the calculation using this value for desired free SO2 level if required Enter this value in the Desired level of free SO2 box Press Calcula...

Page 11: ...ulated I m getting strange results in SO2 mode how do I know if my instrument is working correctly For SO2 measurements with the SC 100A there are several quick tests you can do to make sure the instr...

Page 12: ...connector to its outer metal sheath The device should indicate STOP with its red LED and buzzer or beeper and the current should go to 1999 If this does not happen there may be a problem with the ins...

Reviews: