12.

 

The free SO

2

 content is calculated in units of parts per million (ppm) or mg/L as 

 

ppm SO

2

  =  

64 * V * N * 1000

 

 

 

 

 

 

 

 

 

 

 

 

 

Where V = mL Titrant used to reach the endpoint; N = normality of the Titrant, and S = mL 
wine sample

1

.  If you use 25 mL of sample as directed, and the Titrant is 0.0156 N as supplied, 

then the calculation is simply 

 

 

 

 

ppm Free SO

= 20 * V    

(i.e. 20 times V) 

 

 

Measuring Total SO

2

 by Titration (optional - requires 1N NaOH): 

 

1.

 

Place 25 mL wine or must in the titration beaker (See Figure 5). 

 
2.

 

Add 10ml 1N sodium hydroxide (Vinmetrica Part number SC-100-17) and mix well. Let stand 
approximately 10 minutes. 

 
3.

 

Using the transfer pipettes, add approximately 8 mL of the Acid Reagent and 2 mL of the Reactant 
solution to the vessel. Remember that if you are using the transfer pipettes in the kit, 2 mL is 
marked on the side of the neck of the transfer pipette, so dispense four of these for the Acid 
Reagent. 

 
4.

 

Proceed from step 5 in the Free SO

2

 procedure above. The result calculated will be total SO

2

rather than free SO

2

 in parts per million (ppm) or mg/L. 

 
Finishing up: 
 

1.

 

Turn off the instrument and unplug the electrode. 

 

2.

 

Rinse the SO

electrode and syringe with distilled water.  Let air dry. 

 
3.

 

Store all reagents tightly capped and away from heat and sunlight. 

 
4.

 

Discard waste samples and solutions in accordance with local regulations.  Acidic solutions can be 
neutralized by slow addition of baking soda (sodium bicarbonate) with stirring until effervescence 
ceases.  

                                                 

1

 The complete equation is:  

64 [mg SO

2

/mmol SO

2

 ]* V [mL] * N [meq/mL] * 1000 [mL/L]

 

 

  

 

 

2 [meq/mmol SO

2

] * S [mL]

 

 

2 * S 

 

Summary of Contents for SC-100A

Page 1: ...ge 5 Two 3 mL polyethylene transfer pipettes 6 One 25 mL serological pipette 7 One 100 mL polypropylene beaker Things you will need 1 Two standard AA batteries alkaline type 2 Distilled water which us...

Page 2: ...63 79 99 125 pH 3 0 3 1 3 2 3 3 3 4 3 5 3 6 3 7 3 8 3 9 4 0 We recommend using a sulfite calculator for determining how much sulfite to add to your wines after taking a sulfite measurement with the SC...

Page 3: ...100A shuts off after 30 minutes If this happens unexpectedly just press the POWER button to resume from where you were 4 The SO2 electrode is attached via the BNC connector protruding from the top To...

Page 4: ...to dispense the SO2 Titrant into the top of the burette 3 Using the 25 mL pipette provided in the kit place 25 mL of wine or must in the titration beaker Figure 5 Draw sample up to the 0 mL mark then...

Page 5: ...lectrode Holder adjust it to a similar level You may add up to 20 mL water if needed 8 Verify that the current is less than 50 and the green PROCEED LED is lit If the current is greater than this and...

Page 6: ...gent and 2 mL of the Reactant solution to the vessel Remember that if you are using the transfer pipettes in the kit 2 mL is marked on the side of the neck of the transfer pipette so dispense four of...

Page 7: ...METRICA S MAXIMUM LIABILITY FOR ALL DIRECT DAMAGES INCLUDING WITHOUT LIMITATION CONTRACT DAMAGES AND DAMAGES FOR INJURIES TO PERSONS OR PROPERTY WHETHER ARISING FROM VINMETRICA S BREACH OF THESE TERMS...

Page 8: ...st Mode hold the POWER button down 5 10 seconds until the instrument shuts off If the device does not shut off after 10 seconds of holding down the button move to the next section by releasing then pr...

Page 9: ...2 pH CAL values None only firmware v3 0 6 and later NA Displays current CAL values for pH 7 and 4 in mV Pressing ENTER cycles between these If a pH CAL reset has been done displays CAL value for pH 3...

Page 10: ...r this type of wine molecular SO2 pH is your value mg L Redo the calculation using this value for desired free SO2 level if required Enter this value in the Desired level of free SO2 box Press Calcula...

Page 11: ...ulated I m getting strange results in SO2 mode how do I know if my instrument is working correctly For SO2 measurements with the SC 100A there are several quick tests you can do to make sure the instr...

Page 12: ...connector to its outer metal sheath The device should indicate STOP with its red LED and buzzer or beeper and the current should go to 1999 If this does not happen there may be a problem with the ins...

Reviews: