vinmetrica SC-100A User Manual Download Page 2

SC-100A manual 

Version 3.1c 

Why Test for SO

2

 

 

Testing for sulfite (SO

2

) is crucially important for making sure your wine does not spoil by 

oxidation or from microbial growth. By monitoring your SO

2

 levels, you can make adjustments when 

needed, especially before starting primary fermentation, after malolactic fermentation has completed,  
after racking or when ready to bottle. To correctly adjust sulfite, you need values for your current free 
SO

2

 level and your wine's pH. The SC-100A analyzer will help measure your current free SO

2

. You 

will need to obtain a value for your pH.   
 
 

The key parameter in protecting your wine is molecular SO

2

 which for most wines should be at 

0.5 to 0.8 ppm (mg/L) following secondary fermentation. This in turn depends on the free SO

2

 (it can 

also be referred to as "unbound" SO

2

) and the pH. Overall, you can reach your target molecular SO

2

 by 

measuring and adjusting your free SO

2

 levels and considering your wine's particular pH. See Table 1. 

 

Table

 

1

. Free 

SO

2

 concentrations necessary to attain 0.8 mg/L molecular 

SO

2  

at a designated  pH. 

 

Free 

SO

2

 (ppm) 

13 

16 

21 

26 

32 

40 

50 

63 

79 

99 

125 

pH 

3.0 

3.1 

3.2 

3.3 

3.4 

3.5 

3.6 

3.7 

3.8 

3.9 

4.0 

 
 
 

We recommend using a sulfite calculator for determining how much sulfite to add to your wines 

after taking a sulfite measurement with the SC-100A SO

2

 Analyzer. Winemaker Magazine's Sulfite 

Calculator a

https://winemakermag.com/1301-sulfite-calculator

 can walk you through the process. See 

Appendix B for more information on how to adjust your wine for sulfite. 
 

 

Theory of operation: 

Sulfite (SO

2

):  The SC-100A, with the SO

electrode and reagents provided, can be used to determine 

sulfite (or SO

2

)

  

levels in wine, musts, and other samples.  It relies on the Ripper titration based on the 

quantitative reaction of the SO

2

 with iodine (generated during the titration) which oxidizes the SO

2

 in 

the sample under acid conditions. 

          

 

 

SO

2

  + I

2

 + H

2

 2I

-

 + SO

+ 2H

+

 

  reaction of SO

2

 and iodine 

When all the SO

2

 is titrated at the endpoint, excess iodine appears in solution.  This is detected as 

current with the SO

2

 electrode and signaled by audible and visual indicators. The endpoint is much 

more sensitive than the starch color change commonly employed for Ripper titration, and it is sharp 
and clear, even when titrating red wines and musts.  From the known concentration of the titrant and its 
volume required to reach the endpoint, the free SO

2

 is simply calculated. 

 

 

Summary of Contents for SC-100A

Page 1: ...ge 5 Two 3 mL polyethylene transfer pipettes 6 One 25 mL serological pipette 7 One 100 mL polypropylene beaker Things you will need 1 Two standard AA batteries alkaline type 2 Distilled water which us...

Page 2: ...63 79 99 125 pH 3 0 3 1 3 2 3 3 3 4 3 5 3 6 3 7 3 8 3 9 4 0 We recommend using a sulfite calculator for determining how much sulfite to add to your wines after taking a sulfite measurement with the SC...

Page 3: ...100A shuts off after 30 minutes If this happens unexpectedly just press the POWER button to resume from where you were 4 The SO2 electrode is attached via the BNC connector protruding from the top To...

Page 4: ...to dispense the SO2 Titrant into the top of the burette 3 Using the 25 mL pipette provided in the kit place 25 mL of wine or must in the titration beaker Figure 5 Draw sample up to the 0 mL mark then...

Page 5: ...lectrode Holder adjust it to a similar level You may add up to 20 mL water if needed 8 Verify that the current is less than 50 and the green PROCEED LED is lit If the current is greater than this and...

Page 6: ...gent and 2 mL of the Reactant solution to the vessel Remember that if you are using the transfer pipettes in the kit 2 mL is marked on the side of the neck of the transfer pipette so dispense four of...

Page 7: ...METRICA S MAXIMUM LIABILITY FOR ALL DIRECT DAMAGES INCLUDING WITHOUT LIMITATION CONTRACT DAMAGES AND DAMAGES FOR INJURIES TO PERSONS OR PROPERTY WHETHER ARISING FROM VINMETRICA S BREACH OF THESE TERMS...

Page 8: ...st Mode hold the POWER button down 5 10 seconds until the instrument shuts off If the device does not shut off after 10 seconds of holding down the button move to the next section by releasing then pr...

Page 9: ...2 pH CAL values None only firmware v3 0 6 and later NA Displays current CAL values for pH 7 and 4 in mV Pressing ENTER cycles between these If a pH CAL reset has been done displays CAL value for pH 3...

Page 10: ...r this type of wine molecular SO2 pH is your value mg L Redo the calculation using this value for desired free SO2 level if required Enter this value in the Desired level of free SO2 box Press Calcula...

Page 11: ...ulated I m getting strange results in SO2 mode how do I know if my instrument is working correctly For SO2 measurements with the SC 100A there are several quick tests you can do to make sure the instr...

Page 12: ...connector to its outer metal sheath The device should indicate STOP with its red LED and buzzer or beeper and the current should go to 1999 If this does not happen there may be a problem with the ins...

Reviews: