Appendix A - Test Mode 

 
Test Mode provides various special functions that may be useful in testing the device, for example, if 
troubleshooting is necessary.   
 

 

Test Mode is entered when the POWER button is pressed longer than 2 seconds while turning the 
instrument on.  Remove the pH electrode if it is attached. 

 

Test Mode is organized into sections.  Press the POWER button briefly to move to the next section.  
After the last section, Test Mode restarts the first.  

 

Combinations  of  the  yellow  MODE  LEDs  are  illuminated  to  indicate  the  section  number  as 
shown in the table below.   

 

The Stop LED (red) is illuminated when an error is detected by the instrument  The Proceed LED 
(green) is illuminated to indicate no error detected.  The green LED does not guarantee proper 
functioning; it only indicates that no problem could be automatically detected.  The user should 
make careful observations to discern proper operation. 

 

To exit Test Mode, hold the POWER button down (5-10 seconds) until the instrument shuts off. 
If the device does not shut off after 10 seconds of holding down the button, move to the next 
section by releasing, then pressing again the POWER button briefly; then try to exit again. 

 

Yello

LED

Section 

Equipment Required 

Description 

 

1.  Version 

None. 

The version number of the instrument 
software is displayed. 

 

 
2. Burn-in 

 
None 

The instrument goes through a 
continuous “burn-in” cycle, exercising 
relay, sound, LEDs, and display. 

 

3.  pH 

pH probe or precision 
voltage source.  Do 
not /- 0.5 V. 

[NA] 

An uncalibrated pH level is shown 

in two alternating parts.  First, the integer 
portion of pH level is shown (1 to 14).  
Next, three decimal places are shown.  
Readings above 14.000 are shown as "

---

".  

Readings below 0.000 are shown as 
 " ___".  

 

 

4.  SO

2

 

SO

2

 probe or SO

2

 

probe simulator ( e.g., 
500 kOhm resistor) 

The SO

2

 current in nanoamperes is 

displayed.  For values under 10, one 
decimal place is shown. 

 

5. pH 
Voltage 

pH probe or precision 
voltage source.  Do 
not /- 0.5 V. 

[NA] 

The raw voltage output from the 

instrument's pH amplifier is displayed as 
X.XX volts. Readings can range from 
0.00 to 4.10.

 

Summary of Contents for SC-100A

Page 1: ...ge 5 Two 3 mL polyethylene transfer pipettes 6 One 25 mL serological pipette 7 One 100 mL polypropylene beaker Things you will need 1 Two standard AA batteries alkaline type 2 Distilled water which us...

Page 2: ...63 79 99 125 pH 3 0 3 1 3 2 3 3 3 4 3 5 3 6 3 7 3 8 3 9 4 0 We recommend using a sulfite calculator for determining how much sulfite to add to your wines after taking a sulfite measurement with the SC...

Page 3: ...100A shuts off after 30 minutes If this happens unexpectedly just press the POWER button to resume from where you were 4 The SO2 electrode is attached via the BNC connector protruding from the top To...

Page 4: ...to dispense the SO2 Titrant into the top of the burette 3 Using the 25 mL pipette provided in the kit place 25 mL of wine or must in the titration beaker Figure 5 Draw sample up to the 0 mL mark then...

Page 5: ...lectrode Holder adjust it to a similar level You may add up to 20 mL water if needed 8 Verify that the current is less than 50 and the green PROCEED LED is lit If the current is greater than this and...

Page 6: ...gent and 2 mL of the Reactant solution to the vessel Remember that if you are using the transfer pipettes in the kit 2 mL is marked on the side of the neck of the transfer pipette so dispense four of...

Page 7: ...METRICA S MAXIMUM LIABILITY FOR ALL DIRECT DAMAGES INCLUDING WITHOUT LIMITATION CONTRACT DAMAGES AND DAMAGES FOR INJURIES TO PERSONS OR PROPERTY WHETHER ARISING FROM VINMETRICA S BREACH OF THESE TERMS...

Page 8: ...st Mode hold the POWER button down 5 10 seconds until the instrument shuts off If the device does not shut off after 10 seconds of holding down the button move to the next section by releasing then pr...

Page 9: ...2 pH CAL values None only firmware v3 0 6 and later NA Displays current CAL values for pH 7 and 4 in mV Pressing ENTER cycles between these If a pH CAL reset has been done displays CAL value for pH 3...

Page 10: ...r this type of wine molecular SO2 pH is your value mg L Redo the calculation using this value for desired free SO2 level if required Enter this value in the Desired level of free SO2 box Press Calcula...

Page 11: ...ulated I m getting strange results in SO2 mode how do I know if my instrument is working correctly For SO2 measurements with the SC 100A there are several quick tests you can do to make sure the instr...

Page 12: ...connector to its outer metal sheath The device should indicate STOP with its red LED and buzzer or beeper and the current should go to 1999 If this does not happen there may be a problem with the ins...

Reviews: