Setup

 

Setting up the SC-100A for the first time: 
 

1.

 

The SC-100A (Figure 1) runs on two standard AA batteries (alkaline cells recommended). To 
insert the batteries, open the battery housing on the bottom of the back of the unit by removing the 
two screws and gently prying off the lid.  Install the batteries, then close the housing.  If desired, 
you can prop the unit up using its folding stand. 
 

2.

 

Low Battery Detection: When the battery level is getting low, the instrument shows a low battery 
icon on the upper left side of the display but continues to operate without impairment to any 
function. Replace the batteries as soon as practicable. When the battery level drops too far, the 
instrument does not operate.  It rapidly flashes the low battery icon for 3.0 seconds, beeps and 
shuts itself off. 

 

3.

 

Auto Shut-off:  The SC-100A shuts off after 30 minutes.  If this happens unexpectedly, just press 
the POWER button to resume from where you were. 

 

4.

 

The SO

electrode is attached via the BNC connector protruding from the top. To ensure proper 

function, ensure the BNC is tightened around the SO2 electrode termination. (Figures 2 & 3). 

 

  
 
 
 

5.

 

SO

2

 Electrode: Remove any protective cover from the electrode tip (most electrodes are shipped 

without one). Put the electrode on its side, or hang it from an electrode stand if you have one. The 
SO

2  

electrode is sturdy with its plastic housing, but do take care not to let things touch or strike 

the platinum wires; they are somewhat fragile and will break if bent and straightened repeatedly. 

Electrode care:

 When done, always rinse with DI water and let air dry.  There is no need to store 

any electrode in any kind of solution, and the protective cap should not be replaced. 

 
 
 

Figure 2.

 Attach the SO

2

 

electrode to the BNC connector 

on the SC-100A. 

Figure 3.

 Be sure the SO

2

 

electrode attachment is secure 

by twisting and securing it into 

place. 

Summary of Contents for SC-100A

Page 1: ...ge 5 Two 3 mL polyethylene transfer pipettes 6 One 25 mL serological pipette 7 One 100 mL polypropylene beaker Things you will need 1 Two standard AA batteries alkaline type 2 Distilled water which us...

Page 2: ...63 79 99 125 pH 3 0 3 1 3 2 3 3 3 4 3 5 3 6 3 7 3 8 3 9 4 0 We recommend using a sulfite calculator for determining how much sulfite to add to your wines after taking a sulfite measurement with the SC...

Page 3: ...100A shuts off after 30 minutes If this happens unexpectedly just press the POWER button to resume from where you were 4 The SO2 electrode is attached via the BNC connector protruding from the top To...

Page 4: ...to dispense the SO2 Titrant into the top of the burette 3 Using the 25 mL pipette provided in the kit place 25 mL of wine or must in the titration beaker Figure 5 Draw sample up to the 0 mL mark then...

Page 5: ...lectrode Holder adjust it to a similar level You may add up to 20 mL water if needed 8 Verify that the current is less than 50 and the green PROCEED LED is lit If the current is greater than this and...

Page 6: ...gent and 2 mL of the Reactant solution to the vessel Remember that if you are using the transfer pipettes in the kit 2 mL is marked on the side of the neck of the transfer pipette so dispense four of...

Page 7: ...METRICA S MAXIMUM LIABILITY FOR ALL DIRECT DAMAGES INCLUDING WITHOUT LIMITATION CONTRACT DAMAGES AND DAMAGES FOR INJURIES TO PERSONS OR PROPERTY WHETHER ARISING FROM VINMETRICA S BREACH OF THESE TERMS...

Page 8: ...st Mode hold the POWER button down 5 10 seconds until the instrument shuts off If the device does not shut off after 10 seconds of holding down the button move to the next section by releasing then pr...

Page 9: ...2 pH CAL values None only firmware v3 0 6 and later NA Displays current CAL values for pH 7 and 4 in mV Pressing ENTER cycles between these If a pH CAL reset has been done displays CAL value for pH 3...

Page 10: ...r this type of wine molecular SO2 pH is your value mg L Redo the calculation using this value for desired free SO2 level if required Enter this value in the Desired level of free SO2 box Press Calcula...

Page 11: ...ulated I m getting strange results in SO2 mode how do I know if my instrument is working correctly For SO2 measurements with the SC 100A there are several quick tests you can do to make sure the instr...

Page 12: ...connector to its outer metal sheath The device should indicate STOP with its red LED and buzzer or beeper and the current should go to 1999 If this does not happen there may be a problem with the ins...

Reviews: