highlight to the Exit Door icon
. Press the “OK”
button to select.
3.5 Use the right arrow button to move the black
highlight to the Measure icon.
4. MEASURE MISALIGNMENT
The sensors can be at ANY clock position to start.
The starting measurement is registered by pressing the
“OK” button.
The sensors are then rotated out of the black zone
from the starting point.
The “OK” button registers the second and third
measurements.
The system will give results for the both Vertical and
Horizontal misalignment.
After you get some experience you can try different
starting positions.
To begin, use the orientation shown below.
4.1 Rotate the sensors to 9:00.
Use the animated rotation guide on the display
unit screen.
You should see both “S” and “M” values in the
boxes at the top of the display.
If you do not, then the lasers are not within the
sensors.
4.2 Press the “OK” button to register the 1st
measurement.
The “S” and “M” values are zeroed.
The values are displayed in mils.
Do not rotate shafts while the Measuring icon is
displayed.
4.3 Rotate the sensors to approx. 12:00.
Use the animated rotation guide on the screen.
You should see both “S” and “M” values in the
boxes at the top of the display.
4.4 Press the “OK” button to register the 2nd
measurement.
The “S” and “M” values will not change.
Do not rotate shafts while the Measuring icon is
displayed.
4.5 Rotate the sensors to 3:00.
U S I N G T H E F I X T U R L A S E R G O
10