*1 You can select whether you want to output data in English or in Japanese (katakana) in
Printed Language of Function 1 [
].
*2 Date format depends on the setting of Date Display in Function 1 [
].
12.5.6
Span adjustment/test command
Command Main Body
1st
character
2nd
character
Code
Description
Value
Response
C
0 43H
30H
Disable the Cal key.
*1
C
1 43H
31H
Span adjustment with built-
in weight
(Auto-span adjustment)
C
2 43H
32H
Span test with built-in
weight
C
3 43H
33H
Span adjustment with
external weight
C
4 43H
34H
Span test with built-in
weight
None
A00: Successful
completion
E01: Command
error
E02: Operation is
disabled.
E03: Cancelled by
operation
E04: Abnormal
completion
* This command takes time because the balance sends back a response after an appropriate
operation is completed.
* If the Cal key is disabled in Span Adjustment/Span Test of Function 1 [
], a span
adjustment/test command does not work.
* If the Cal key is disabled, however, the setting is reset to the function setting when the balance
is turned on again.
12.5.7
Sample input commands
T(SP)(CR)(LF)
Set a tare weight or adjust the zero point.
O1(CR)(LF)
Set the balance to continuous output.
O8(CR)(LF)
Output data (once immediately).
IA,12,34,56(CR)(LF)
Set an interval of 12 hrs. 34 min. 56 sec.
OA(CR)(LF)
Start the interval function.
DD(CR)(LF)
Output the date.
DT(CR)(LF)
Output the time.
C1(CR)(LF)
Execute span adjustment with the built-in weight.
53