8.5
Judge by Deviation Values
8.5.1
Set two limit values by putting actual samples on the balance
1 Start the limit function.
Press the Set key for a few seconds.
When a series of dots [
] is displayed, release
the key.
The current reference weight blinks in the display.
2 Save the reference weight.
Put a sample that provides a reference weight on the
balance, and press the Function key.
When a reference weight is saved, that value is
displayed for a while.
3 Save a lower limit.
The display first indicates [
], and then the
current lower limit blinks.
Put a sample that provides a lower limit on the
balance, and press the Function key.
A lower limit is saved. Then go to the next step.
4 Save an upper limit.
The display first indicates [
], and then the
current upper limit blinks.
Put a sample that provides an upper limit on the
balance, and press the Function key.
The upper limit is saved, and the balance goes back
into measurement mode.
Continuous press
* To set three or four points, repeat steps 3 through 4 above.
Saved limit values are displayed as [
] - [
] or [
] rather than
[
] or [
]. In this case, a bar graph is displayed to indicate a rank.