Page 36 of 43
5.6.4 RF Distance to Fault (DTF)
The following procedure describes the steps to perform distance to fault tests (Return Loss or VSWR) with
OneAdvisor.
Step Action
Description
1
DTF measurement mode:
-
Select the desired
measurement layout.
-
Select the corresponding
DTF measurement icon (RTF
in Return Loss or DTF in
VSWR).
Note: Refer to the “Initial Setup”
and “RF Reflection Test” sections
for initial configuration,
connectivity and reflection test.
DTF Measurement Types
2
Configure the DTF measurement:
-
Select the configuration icon
and select {General}
-
Set the desired Data Points,
Interference Rejection,
Windowing, Units, and Bias.
General Cable and Antenna Settings
3
Configure the DTF distance
measurement:
-
Select the measurement title
{General}
-
Select {Distance}
-
Set the desired Start
Distance, and Stop Distance.