Fiber Type /
Connector
Tip(s) to use on
FiberChek/P5000i
Cleaning options
BAP-3 over the fibers. See
right >>
SC-APC
FBPT-SC-APC OTDR Test Port
SC-APC
Connector on
OTDR Launch
Fiber
FBPT-U25MA
Note that the TBERD 5800 OTDR Port and the 10M launch
fiber may have the SC-APC style green connection. A is for
Angled. The 8-degree angle on the end face creates a
superior connection in terms of reflectance and durability.
The OTDR port can be cleaned with a SC style IBC stick
cleaner (shown above). The fiber itself can be cleaned with
that as well or with wipes, CLETOP, etc. The tips for the
inspection scope have a built-in angle as well. Be sure and
inspect and then protect the cleanliness of your OTDR port
and launch fiber.
5.1.5 Fiber Inspection Testing – p5000i
Step Action
Description/Diagram
1
Press and hold the ON/OFF
button to turn on the ONA-800.
Press the Home Key
Select the Microscope Tab