Note: To compensate for the influence of CO
2
on the
N
2
O it is important the Zero and
N
2
O Span calibrations do not contain CO
2
.
CO Cell:
Carbon monoxide is measured by an electrochemical gas sensor. It is not affected by the
other measurement gases but can have cross interference from gases not normally present
in medical supply gases and suffers virtually no influence from CO
2
, H
2
S, NO
2
, SO
2
.
8.6 Hardware Reset
If for any reason the instrument ‘locks up’ and will not switch off, it is possible to force a
hardware reset. Press and hold the ‘On/Off’ key for ten seconds; wait at least 15 seconds
and the instrument should restart automatically.
?
Note: Performing a hardware reset may cause loss or corruption of currently stored data
including the time/date.