5.12.2.5 Flow Fail
The ‘Flow Fail’ option enables the operator to adjust the instrument flow fail detection
point should it fail in normal operation with a clean filter.
The instrument’s internal pump can be stalled when pulling against a vacuum or through
a blocked filter. This is indicated by a flashing pump icon ; to prevent damage to the
pump, the pump will switch off after a few seconds. Press the ‘Pump’ key again to
remove the flashing pump icon.
?
Note: Dirty or discolored filters should be changed before use. Filters that have
drawn in water should be changed immediately to prevent damage to the
instrument.
1)
From the ‘Main Read Screen’ press the ‘Menu’ key on the instrument panel.
2)
Press ‘Key 2’ to display the ‘Utilities’ menu.
3)
Press ‘Key’ 5 to select ‘Flow Fail’ and the following screen is displayed:
Adjust Flow Fail
4)
Use the scroll keys ‘Key 4 – Scroll Left’ or ‘Key 6 – Scroll Right’ to adjust the value
displayed. The larger the value, the less sensitive the flow fail detection is.
5)
Press soft-key ‘Accept’ or ‘Reject’ accordingly.
For more information please refer to the following Flow Fail Process.