6.3 G210 Gas Measurement Process – Best Practice
1)
When the instrument is first switched on it should be purged with fresh air and allowed
to stabilize for a few minutes.
2)
Make sure that you have carried out all the preliminary checks as listed in section ‘6.1
Preliminary Checks – Best Practice’ and that the instrument is ready to take the first
reading. Always use a regulator pressure kit with the G210 instrument.
3)
Connect the
sample tube, if relevant, from the sample point to
the inlet
port of the
instrument, ensuring the filter is seated correctly.
4)
If auto-logging is required select the ID codes.
5)
If auto-logging is not required, press the pump key to draw a sample, stop the pump
and press the soft-key ‘Store’ to store the reading. The operator will then be prompted
to enter or select ID codes. Select ID codes and press the ‘Accept’ button to confirm.
A message will display confirming that the reading has been stored.
6)
The instrument automatically takes the operator back to the ‘Main Read Screen’ ready
to take another reading.
Gas Measurement Process
(G210 - Best Practice)
(Switch on)
Instrument
warm up
Main Read
Screen
Run a
Clean Air
Purge
Connect
Sample Tube
if required
Start
Reading Stored
(confirmation
message
displayed)
Enter ID
Code
& press
Enter
END
Press the
‘Store’ key
Disconnect
the Sample
Tube from
the inlet
port
Instrument automatically
returns to the
‘Main Read Screen’
Run a
Clean Air
Purge
Make sure
that you
have
carried out
all the
preliminary
checks prior
to taking a
reading
Is Auto-logging
required?
Set ID &
Logging
parameters
NO
YES
Readings are
taken
automatically
Start
Logging
Take another
reading?
Stop
Logging
Gas
measurement is
taken
automatically
Wait for Main
Gas Read
Screen to
stabilise
Take another
reading?
YES
NO
NO
YES