OMG200N1.24
G200 Analyzer Range
VIASENSOR
Page
18
will last 8-10 hours.
When the instrument is already powered on, the operation is slightly different as the battery
icon changes to a flashing plug symbol. This will stop flashing when the charge is complete.
5.10 Instrument Main Read Screen
After the analyzer has been switched on and the warm-up self-tests completed
(Approximately 45 minutes) the analyzer will display the following screen.
Main Read Screen
Soft-Keys:
Peak - Enables the operator to display the peak reading.
Store - Enables the operator to store the reading for viewing/download later.
Hold - Enables the operator to hold the current reading being taken.
5.10.1 Alternative Reading Methods
The G210 has three reading types or methods which each require slightly different
operating procedures:
•
Peak reading
•
Store reading
•
Hold reading
Peak Reading:
The operator can toggle the reading mode between normal (current) and peak readings.
While in peak reading mode the instrument will only display peak values for each of the
channels. These values can then be stored by pressing the ‘Store’ key or automatically at
the appropriate logging interval (if logging is enabled).
The peak value is reset after a reading is stored or by exiting the peak mode using the
appropriate soft-key. The current mode of operation can be identified by the status of the
soft-key, either ‘Normal’ or ‘Inverse’, where inverse indicates peak mode is active.
Store Reading:
The ‘Store’ reading option allows the operator to store the current reading. Upon selecting
this option the operator is then prompted to enter or select a ‘Site ID’ and ‘Sample Point
ID’ to identify the reading. Select the ‘Accept’ button to confirm the ID choices. For
further information please refer to section ‘5.11 Enter ID Codes’.
Hold Reading: