Manual page 7
SETUP MENU DESCRIPTIONS AND PROCEDURES
Factory-set defaults are shown in
bold
.
NAME/CODE
DESCRIPTION
CODE/VALUE
F1
Graduations
Specifies number of full-scale graduations. Value should be consistent with legal
requirements and environmental limits on the useful system resolution.
500 1,000
1,500 2,000
2,500 3,000
4,000
5,000
6,000 8,000
10,000 12,000
20,000 30,000 40,000 50,000
F2
Span Gain
Span Gain is related to A/D integration time. The larger the span gain, the higher the
internal resolution, but the slower the update speed. Note that the scale must be re-
calibrated whenever this parameter is altered. See Appendix C for more information.
25 50
75
100
150 200
F3
Zero Track Band
Selects the range within which the scale will automatically zero. Note that the scale
must be in standstill to automatically zero. Selections are in Display Divisions.
0d
0.5d
1d 3d 5d
F4
Zero Range
Selects the range within which the scale may be zeroed. Note that the indicator must
be in standstill to zero the scale.
100% (US)
1.9% (CE)
2%
F5
Motion Band
Sets the level at which motion is detected by comparing the present display update
with the previous one. If motion is not detected for two seconds or more, scale is in
standstill and can process a Print or Zero command. Maximum value varies depending
on local regulations.
0.25d (CE) 1d (US)
3d
5d 10d
F6
Digital Filter
Averages weight readings to produce higher stability. The higher the filter setting, the
greater the stability but the slower the indicator’s response time. Choose 8 unless a
very fast response is needed.
1 2
4
8
F7
Overload Limit
Selects the desired formula which determines the point at which the indicator shows
overload. All selections are based on the primary unit selected in F8. "FS" = Full scale
in primary units.
FS
FS + 2% (US)
FS + 1d
FS + 9d (CE)
F8
Calib. Unit
Selects the primary base unit to be used in the calibration process. Also the default unit
for normal operation.
"1" = primary unit is lb. "2" = primary unit is in kg.
1 (US)
2 (CE)
F9
Display
Divisions
Determines the desired weight increments. Value should be consistent with legal
requirements.
1
2 5
F10
Decimal Pt.
Determines location of the decimal point.
0
0.0
0.00 0.000
0.0000 00
F16
Zero Calibration
Places indicator into the zero calibration routine. Scrolling down with the
ZERO
key
one level begins the procedure.
Press
ZERO
key to begin
sequence
F17
Span Calibration
Places indicator into the span calibration routine. Scrolling down with the
ZERO
key
one level begins the procedure.
Press
ZERO
key to begin
sequence
F18
View Calibration
Actuates the function that allows you to view both the zero and span calibration value.
The values displayed in this function are valid only after Calibration (F16 & F17) has
been successfully completed. Scrolling down with the
ZERO
key one level begins the
procedure.
Press
ZERO
key to begin
sequence
F19
Key-in Zero
Allows you to key-in known zero calibration value in case of memory loss in the field.
Scrolling down with the
ZERO
key one level begins the procedure.
Press
ZERO
key to begin
sequence
F20
Key-in Span
Allows you to key-in a known span calibration value in case of memory loss in the
field. Scrolling down with the
ZERO
key one level begins the procedure.
Press
ZERO
key to begin
sequence
F21
Factory Reset
This sub-menu will reset all parameters in the “F” and “A” menu to the default
settings. USE WITH CAUTION!
Press the
ZERO
key twice to
execute.