KEY-IN ZERO CALIBRATION VALUE (F19)
Note:
This procedure is intended for emergency use only in the case of non-volatile memory loss. A valid zero calibration value,
obtained from a successful F16 calibration procedure, must be used.
1. While in the Setup mode, scroll to "
F 19
", then scroll down once using the
ZERO
key.
2. The display will momentarily show "
CAL 0
", followed by a flashing zero. Use the four directional keys to adjust the
displayed value to the zero calibration value.
3. After setting the exact value, press the
NET/GROSS
key to save the value.
4. The display will show "
E CAL 0
" momentarily, then revert back up to F19.
KEY-IN SPAN CALIBRATION VALUE (F20)
Note:
This procedure is intended for emergency use only in the case of non-volatile memory loss. A valid span calibration value,
obtained from a successful F17 calibration procedure, must be used.
1. While in the Setup mode, scroll to "
F 20
", then scroll down once using the ZERO key.
2. The display will momentarily show "
CAL 1
", followed by a flashing zero. Use the four directional keys to adjust the
displayed value to the span calibration value.
4. After setting the exact value, press the
NET/GROSS
key to save the value.
5.
If the entered value is greater than zero, the display will show "
E CAL 1
" momentarily, then revert back up to F20. If a
value of zero is entered, the indicator will briefly show "
Err 5
", then revert back to the screen described above in Step # 2.