background image

Manual   page 11 

CALIBRATION 

The indicator is calibrated by following the procedures embedded in F16 (Zero) and F17 (Span) of the Setup Menu.  
Each procedure enters a value into the indicator's non-volatile memory - F16 the zero value (deadweight) and F17 the span value (test weight). 
The minimum test weight that can be used is 1% of full-scale capacity. After the two calibration procedures are executed successfully, you 
should record both calibration values using the F18 View procedure. 

In the unlikely event that either value is lost while in the field, the setup menu makes provisions for re-entering these values via F19 and F20, 
thus eliminating the need for re-calibration with test weights. 

NOTE:

 This chapter assumes that the indicator is in Setup (“F”) Menu mode. If the indicator is not in Setup Menu mode, refer to Chapter 3 for 

instructions. 

ZERO CALIBRATION (F16)

 

1. While in the Setup mode, scroll to "

F 16

", then scroll down once using the 

ZERO

 key to enter zero calibration menu. The display will

momentarily show "

C 0

" followed by a value. This value is the internal A/D count and can prove useful when trying to troubleshoot setup

problems.

2. After making sure that there are no test weights on the platform, press the 

ZERO

 key to zero out the displayed value.

3. Press the 

NET/GROSS

 key to save the zero point value. The display will show "

EndC0

" momentarily, then revert back up to F16. At this

time, proceed to the F17 span calibration to complete indicator calibration.

SPAN CALIBRATION (F17)

 

1. While in the Setup mode, scroll to "

F 17

", then scroll down once using the 

ZERO/OFF

 key to enter span calibration menu.

2. The display will momentarily show "

C 1

" for the span calibration, followed by a value with one flashing digit. This value will be zero with

the Decimal Point parameter selected in F10. Place the test weight on the weighing mechanism.

3. Use the four directional keys to adjust the displayed value to the actual test weight value. Increase the flashing digit by pressing the 

lb/kg

key. Decrease the flashing digit by pressing the 

ZERO

 key. Pressing the 

TARE

 key or the 

PRINT

 key will change the position of the

flashing digit.

4. After setting the exact value, press the 

NET/GROSS

 key to save the value.

5. If the calibration was successful, the display will show "

EndC1

" momentarily, then revert back up to F17. At this time it is suggested that

the calibration values be recorded for future use.

6. If the calibration was 

not 

successful, one of the error messages below will appear. Take the indicated action to correct the problem, then

perform a new calibration.

 "

Err0

The calibration test weight or the adjusted keyed-in weight is larger than the full capacity of the scale. Change the 
calibration test weight or check the input data. 

 "

Err1

The calibration test weight or the adjusted keyed-in weight is smaller than 1% of the full capacity of the scale. 
Change the calibration test weight or check the input data. 

"

Err2

The internal resolution of the scale is not high enough to accept the calibration value. Select a larger parameter for 
the Span Gain (F2). 

VIEW CALIBRATION VALUES (F18) 

Note: 

The values displayed in this procedure are valid only after a successful calibration has been performed using F16 and F17. 

1. While in the Setup mode, scroll to "

F 18

", then scroll down once using the ZERO key to enter View calibration menu.

2. The display will momentarily show "

CAL 0

" followed by a value. This value is the 

zero calibration value

 and should be

recorded in the table below. Press any key to continue.

3. The display will momentarily show "

CAL 1

" followed by another value. This value is the 

span calibration value

 and should

also be recorded in the table below. Press any key to return to upper level (F18).

INDICATOR

ZERO 

CALIBRATION 

VALUE 

SPAN CALIBRATION VALUE 

S/N:

Summary of Contents for TI-1200

Page 1: ...geable battery and with a smaller cabinet Note When TI 1200 is shipped as a part of a complete Scale e g AV 250 seies and AL 100 21 the Scales are normally calibrated When TI 1200 is shipped together with Loadcells it is adjusted to the application but a calibration is always needed at the place of installation EC type approved non automatic weighing instrument Number FI 01 1 04 Accuracy class Man...

Page 2: ...erate identically All readout up to 50 000 display divisions on a liquid crystal display LCD with back light All set up parameters may be entered via the front panel keys including calibration An external power supply is included The external power supply functions 9 VDC as a charger for the rechargeable battery The power supply may also be used as the main power supply TEST CERTIFICATE Number FIT...

Page 3: ...pplications is the charger board disconnected and the power is connected direct to J2 In this case the battery normally is removed If you want it is possible to demount the contact and connect the cable direct to battery charger or J2 Note centre contact The cable from the Loadcell s should be connected to J1 Remove a sufficient part of the external isolation to avoid strength on the wires Colours...

Page 4: ...able direct to battery charger or J2 Note centre contact Calibration Normal The cable from the Loadcell s should be connected to J1 Remove a sufficient part of the external isolation to avoid strength on the wires Colours se next page Connect the shield to E Display PCB Calibration switch Automatic start Automatic start Short transistor Q1 between above two points ...

Page 5: ...used as a simplex RS 232 port designed for connection to a remote display Connect your serial printer remote display or computer communication cable not included to terminal J3 on the main board J3 GND RXD TXD nc nc Figure 2 3 Connection assignments for the serial communication terminal Terminal J1 Loadcell Input Output Input Output Loadcell Input Output Input Output If more than one Loadcells all...

Page 6: ...press ZERO OFF down key once The current saved selection is shown 3 To move to a new F heading use TARE left or ON PRINT right key to move right or left in the Setup Menu Chart 4 If you want to go back whithout saving press lb kg 5 If a digit flashing use lb kg and ZERO OFF for choosing value 6 To save a new selection press the NET GROSS Set key To exit without saving press the PRINT up key to ret...

Page 7: ...he desired formula which determines the point at which the indicator shows overload All selections are based on the primary unit selected in F8 FS Full scale in primary units FS FS 2 US FS 1d FS 9d CE F8 Calib Unit Selects the primary base unit to be used in the calibration process Also the default unit for normal operation 1 primary unit is lb 2 primary unit is in kg 1 US 2 CE F9 Display Division...

Page 8: ...o 3 but do not save the zero point 2 After pressing ZERO to zero the offset place the test weight s on the platform The displayed count is the internal count If the count remains on zero check your load cell connections 3 At full scale the displayed count should be a minimum of 2 times the desired external graduations However for maximum stability a ratio of 6 1 or higher is recommended 4 If the d...

Page 9: ...t the power A1 Baud Rate 1200 2400 4800 9600 A2 Data Bits Parity 8n 7O 7E A3 Transmission Mode C d A4 Display Check A5 Enable lb kg Key 0 1 Press ZERO key to begin A7 ID No Enable A6 Serial Port Mode 0 1 0 1 A8 ID No Entry Press ZERO key to begin A9 No of Line Feeds Press ZERO key to begin 7n A10 Auto Power Off 5 8 10 15 OFF 1 2 3 20 30 ID Number Entry A8 1 While in the User Menu mode scroll to A ...

Page 10: ... to scroll down one level begins the test sequence Press ZERO key to begin sequence A5 Disable the lb kg Key Allows the lb kg key to be disabled so that an operator cannot accidentally press the key and change the displayed units 0 Disable the lb kg key 1 Enable the lb kg key 0 CE 1 US A6 Serial Port Mode Selects the mode of the RS 232 serial port Refer to Bilaga B for more information 0 Full Dupl...

Page 11: ...lected in F10 Place the test weight on the weighing mechanism 3 Use the four directional keys to adjust the displayed value to the actual test weight value Increase the flashing digit by pressing the lb kg key Decrease the flashing digit by pressing the ZERO key Pressing the TARE key or the PRINT key will change the position of the flashing digit 4 After setting the exact value press the NET GROSS...

Page 12: ...en revert back up to F19 KEY IN SPAN CALIBRATION VALUE F20 Note This procedure is intended for emergency use only in the case of non volatile memory loss A valid span calibration value obtained from a successful F17 calibration procedure must be used 1 While in the Setup mode scroll to F 20 then scroll down once using the ZERO key 2 The display will momentarily show CAL 1 followed by a flashing ze...

Page 13: ...ROM 256 x 8 external to uC Digital Filtering Software selectable SERIAL COMMUNICATIONS Serial Port Full Duplex 1200 2400 4800 9600 Baud 8 data bits no parity 1 stop bit 7 data bits odd parity 1 stop bit 7 data bits even parity 1 stop bit 7 data bits no parity 2 stop bits OPERATOR INTERFACE Display 0 75 19 mm 7 segment Liquid Crystal 6 Digit Additional Symbols Net Gross Stable Tare lb kg Zero Keybo...

Page 14: ... will not respond if the scale is in motion positive overload or negative overload Z This command is sent to the indicator to zero the scale The indicator will not respond if the scale is in motion positive overload or negative overload The indicator will also not respond if it is not in gross mode or within the zero range specified in F4 of the Setup Menu T This command is sent to the indicator t...

Page 15: ... A7 of the User Menu ID NO 123456 GROSS 25 00 LB TARE 1 48 LB NET 23 52 LB INDICATOR PRINTER 3 RXD CTS 2 TXD 3 S GND 5 20 DTR 7 S GND DSUB9 DSUB25 FIGURE B 3 Print Ticket FIGURE B 4 Cable Diagram for Indicator to Printer SIMPLEX MODE The Simplex Mode provides a continuous serial transmission mode and is selected by setting A3 to C and A6 to 0 The Continuous mode is used to interface to computers s...

Page 16: ...dicator needs service Err 5 Key in Span Calibration Mode F20 You have attempted to enter a zero value for C1 Enter a known calibration value greater than zero Err 7 Initialization No reading from the ADC Make sure there is a load cell s connected to the indicator at start up Err 9 Normal Operating Mode Span calibration value has been lost Re calibrate the scale Fault localization If it is problem ...

Page 17: ...ht is within 0 25 divisions of true zero N Indicates that the indicator is displaying net weight G Indicates that the indicator is displaying gross weight T Indicates that a tare weight has been established in the system lb kg Indicates the unit of the displayed weight Indicates a low battery condition Re charge the battery Indicates stable reading ...

Page 18: ...tor is off the PRINT key is used to turn the indicator on GENERAL SCALE OPERATION WEIGHING AN ITEM 1 Select the desired weighing unit by pressing the lb kg key until that unit is indicated on the display 2 If necessary press the ZERO key to obtain a weight reading of zero 3 Place the object to be weighed on the scale s platter and allow the weight indication to stabilize If the item weight exceeds...

Reviews: