7-2
VL-12CT96/7 Analog & Digital I/O Card
Analog Output Calibration (Voltage Mode)
Required Equipment
• A voltmeter with resolution and accuracy to ½ LSB of the output range being used.
• A small flat blade screwdriver.
Calibration Procedure
Channel 0
• Disconnect cable from connector J2.
• Connect the voltmeter to channel 0 on connector J2; positive lead (red) to pin 2, negative lead
(black) to pin 1.
• Write 0800H to channel 0.
• Adjust Z0 pot until voltage is exactly -10.0000V.
• Write 07FFH to channel 0.
• If using ±5V range, adjust G0 pot until voltage is e4.997559V.
If using ±10V range, adjust G0 pot until voltage is e9.995117V.
Channel 1
• Connect the voltmeter to channel 1 on connector J2; positive lead (red) to pin 8, negative lead
(black) to pin 7.
• Write 0800H to channel 1.
• Adjust Z1 pot until voltage is exactly -10.0000V.
• Write 07FFH to channel 1.
• If using ±5V range, adjust G1 pot until voltage is e4.997559V.
If using ±10V range, adjust G1 pot until voltage is e9.995117V.
Calibration — Analog Output (Voltage Mode)