5.0
Operation
5.1
Operating Cycle
This product is fitted with an instrument switch which cuts off power to the control
circuit.
Connect the product to the electrical supply.
Turn on the instrument switch to activate the temperature controllers. The controllers
illuminate and go through a short test cycle.
Over-Temperature option only
. If the digital over-temperature option has not yet
been set as required, set and activate it according to the over-temperature controller
instructions.
The product will heat up according to the controller setpoint or program, unless a time
switch is fitted and switched off.
As the product heats up, the heat light glows steadily at first and then flashes as the
product approaches the desired temperature. For more information on temperature
control see the controller instructions.
Over-Temperature option only.
If the over-temperature circuit has tripped, an
indicator on the over-temperature controller flashes and the heating elements are
isolated. Find and correct the cause before resetting the over-temperature controller
according to the instructions supplied.
To turn the product off, set the instrument switch to its off position. The controller
display will go blank. If the product is to be left unattended, isolate it from the electrical
supply.
5.2
General Operating Advice
Heating element life is shortened by overheating. Do not leave the
product at high temperature when it is not required. The maximum
temperature is shown on the product rating label and in section 11.0
towards the back of this manual.
Lightweight ceramic fibre insulation can easily be marked by accidental
contact. Some fine cracks may develop in the surface of the insulation
due to the progressive shrinkage of the insulation materials. Cracks are
not usually detrimental to the functioning or the safety of the product.
25
5.0 Operation
Summary of Contents for CARBOLITE GERO VMF-ASTM
Page 4: ...10 2 Power Settings 38 11 0 Specifications 39 11 1 Environment 39 4 ...
Page 20: ...3 0 2132 Controller 3 7 Navigation Diagram 20 ...
Page 28: ...6 0 Maintenance Power Consumption Measure the current drawn on each phase cir cuit 28 ...
Page 40: ......
Page 41: ...Notes Service Record Engineer Name Date Record of Work ...