6.0
Maintenance
6.1
General Maintenance
Preventive rather than reactive maintenance is recommended. The type and frequency
depends on the product use; the following are recommended.
6.2
Maintenance Schedule
CUSTOMER
QUALIFIED PERSONNEL
DANGER! ELECTRIC SHOCK
. Risk of fatal injury. Only electrically
qualified personnel should attempt these maintenance procedures.
Maintenance
Procedure
Method
Frequency
Daily Weekly Monthly
Bi-
Annually
Annually
Safety
Over-Temperature Safety
Circuit
(if fitted)
Set an over-temperature setpoint lower than the
displayed temperature and check for an over-tem-
perature alarm as detailed in this manual
Over-Temperature Safety
Circuit
(if fitted)
Electrical measurement
Lid Insulation
Visual inspection, checking the seal and whether
it is free of damage
Lid Insulation
Replacement where necessary
Electrical Safety (external)
Test the 30mA RCD trip with the test button daily
before use.
Electrical Safety (external)
Visual check of external cables and plugs
Electrical Safety (internal)
Physically check all connections and cleaning of
the power plate area
Function
Temperature Calibration
Tested using certified equipment, frequency
dependent on the standard required
Operational Check
Check that all functions are working normally
Operational Check
Thorough inspection and report incorporating a
test of all functions
Performance
Element Circuit
Electrical measurement
27
6.0 Maintenance
Summary of Contents for CARBOLITE GERO VMF-ASTM
Page 4: ...10 2 Power Settings 38 11 0 Specifications 39 11 1 Environment 39 4 ...
Page 20: ...3 0 2132 Controller 3 7 Navigation Diagram 20 ...
Page 28: ...6 0 Maintenance Power Consumption Measure the current drawn on each phase cir cuit 28 ...
Page 40: ......
Page 41: ...Notes Service Record Engineer Name Date Record of Work ...